DC coupling, RF attenuation ≥10 dB,
preselection = ON, preamplifier = ON
8
+20 °C to +30 °C
1 kHz ≤ f < 10 MHz
<0.8 dB (σ = 0.26 dB)
10 MHz ≤ f < 2 GHz
<0.6 dB (σ = 0.2 dB)
2 GHz ≤ f < 3.6 GHz
<0.8 dB (σ = 0.26 dB)
3.6 GHz ≤ f < 8 GHz, span < 1 GHz
<1,5 dB (σ = 0.7 dB)
8 GHz ≤ f ≤ 40 GHz, span < 1 GHz
<2.0 dB (σ = 0.7 dB)
RF attenuation > 40 dB or
f ≥ 3.6 GHz, span ≥ 1 GHz
add 0.5 dB to above values
+5 °C to +45 °C
1 kHz ≤ f < 3.6 GHz
<0.9 dB (σ = 0.3 dB)
3.6 GHz ≤ f < 22 GHz
add 0.5 dB to above values
22 GHz ≤ f < 26.5 GHz
add 1.0 dB to above values
26.5 GHz ≤ f ≤ 40 GHz
add 2.0 dB to above values
RF attenuation > 40 dB or
f ≥ 3.6 GHz, span ≥ 1 GHz
add 0.5 dB to above values
R&S
®
ESU26, R&S
®
FSU40:
Second harmonic intercept (SHI)
f
in
> 1.8 GHz
>65 dBm, nominal
RF attenuation <10 dB, DC coupled,
+5 °C to +30 °C
f ≤ 1 GHz
<2.0
1 GHz ≤ f < 2.5 GHz
<3.0
2.5 GHz < f ≤ 3.6 GHz
<4.0
VSWR
f ≥ 3.6 GHz
<3.0
Ordering information
Designation Type
Order
No.
EMI Test Receiver 20 Hz to 8 GHz
R&S
®
ESU8 1302.6005.08
EMI Test Receiver 20 Hz to 26.5 GHz
R&S
®
ESU26 1302.6005.26
EMI Test Receiver 20 Hz to 40 GHz
R&S
®
ESU40 1302.6005.40
Accessories supplied
Power cable, probe power cable, printed quick start guide and CD-ROM (with operating manual and service manual)
R&S
®
ESU26: test port adapter with 3.5 mm female (1021.0512.00) and N female (1021.0535.00) connector
R&S
®
ESU40: test port adapter with K female (1036.4790.00) and N female (1036.4777.00) connector
Options
Designation Type
Order
No.
Retrofittable
Remarks
Options
OCXO, low aging/ improved phase noise
at 10 Hz carrier offset
R&S
®
FSU-B4 1144.9000.02 yes
Tracking Generator, 100 kHz to 3.6 GHz
R&S
®
FSU-B9 1142.8994.02 yes
External Generator Control
R&S
®
FSP-B10 1129.7246.03 yes
Output Attenuator, 0 dB to 70 dB, for
R&S
®
FSU-B9
R&S
®
FSU-B12 1142.9349.02 yes
requires
R&S
®
FSU-B9
Removable Hard Disk
R&S
®
ESU-B18 1303.0400.06 no
Second Hard Disk for R&S
®
ESU-B18
R&S
®
ESU-B19 1303.0500.06 –
requires
R&S
®
ESU-B18
Low-Noise Preamplifier
(100 kHz to 8 GHz)
R&S
®
ESU-B24 1157.2100.08 yes
for
R&S
®
ESU8 only
Low-Noise Preamplifier
(100 kHz to 26.5 GHz)
R&S
®
ESU-B24 1157.2100.26 yes
for
R&S
®
ESU26 only
Low-Noise Preamplifier
(100 kHz to 40 GHz)
R&S
®
ESU-B24 1157.2100.40 yes
for
R&S
®
ESU40 only
Firmware/software
Time Domain Scan
R&S
®
ESU-K53 1305.8509.02 yes
key
code
Triggered Scan for Coverage Meas.
R&S
®
ESPIK50 1106.4386.02 yes
key
code
8