22 R&S

®

ESU EMI Test Receiver

Version 02.00, October 2007

DC coupling, RF attenuation ≥10 dB,
preselection = ON, preamplifier = ON

8

+20 °C to +30 °C

1 kHz ≤ f < 10 MHz

<0.8 dB (σ = 0.26 dB)

10 MHz ≤ f < 2 GHz

<0.6 dB (σ = 0.2 dB)

2 GHz ≤ f < 3.6 GHz

<0.8 dB (σ = 0.26 dB)

3.6 GHz ≤ f < 8 GHz, span < 1 GHz

<1,5 dB (σ = 0.7 dB)

8 GHz ≤ f ≤ 40 GHz, span < 1 GHz

<2.0 dB (σ = 0.7 dB)

RF attenuation > 40 dB or
f ≥ 3.6 GHz, span ≥ 1 GHz


add 0.5 dB to above values

+5 °C to +45 °C

1 kHz ≤ f < 3.6 GHz

<0.9 dB (σ = 0.3 dB)

3.6 GHz ≤ f < 22 GHz

add 0.5 dB to above values

22 GHz ≤ f < 26.5 GHz

add 1.0 dB to above values

26.5 GHz ≤ f ≤ 40 GHz

add 2.0 dB to above values

RF attenuation > 40 dB or
f ≥ 3.6 GHz, span ≥ 1 GHz


add 0.5 dB to above values

R&S

®

ESU26, R&S

®

FSU40:

Second harmonic intercept (SHI)

f

in

> 1.8 GHz

>65 dBm, nominal

RF attenuation <10 dB, DC coupled,
+5 °C to +30 °C

f ≤ 1 GHz

<2.0

1 GHz ≤ f < 2.5 GHz

<3.0

2.5 GHz < f ≤ 3.6 GHz

<4.0

VSWR

f ≥ 3.6 GHz

<3.0

Ordering information

Designation Type

Order

No.

EMI Test Receiver 20 Hz to 8 GHz

R&S

®

ESU8 1302.6005.08

EMI Test Receiver 20 Hz to 26.5 GHz

R&S

®

ESU26 1302.6005.26

EMI Test Receiver 20 Hz to 40 GHz

R&S

®

ESU40 1302.6005.40

Accessories supplied
Power cable, probe power cable, printed quick start guide and CD-ROM (with operating manual and service manual)
R&S

®

ESU26: test port adapter with 3.5 mm female (1021.0512.00) and N female (1021.0535.00) connector

R&S

®

ESU40: test port adapter with K female (1036.4790.00) and N female (1036.4777.00) connector

Options

Designation Type

Order

No.

Retrofittable

Remarks

Options
OCXO, low aging/ improved phase noise
at 10 Hz carrier offset

R&S

®

FSU-B4 1144.9000.02 yes

Tracking Generator, 100 kHz to 3.6 GHz

R&S

®

FSU-B9 1142.8994.02 yes

External Generator Control

R&S

®

FSP-B10 1129.7246.03 yes

Output Attenuator, 0 dB to 70 dB, for
R&S

®

FSU-B9

R&S

®

FSU-B12 1142.9349.02 yes

requires

R&S

®

FSU-B9

Removable Hard Disk

R&S

®

ESU-B18 1303.0400.06 no

Second Hard Disk for R&S

®

ESU-B18

R&S

®

ESU-B19 1303.0500.06 –

requires

R&S

®

ESU-B18

Low-Noise Preamplifier
(100 kHz to 8 GHz)

R&S

®

ESU-B24 1157.2100.08 yes

for

R&S

®

ESU8 only

Low-Noise Preamplifier
(100 kHz to 26.5 GHz)

R&S

®

ESU-B24 1157.2100.26 yes

for

R&S

®

ESU26 only

Low-Noise Preamplifier
(100 kHz to 40 GHz)

R&S

®

ESU-B24 1157.2100.40 yes

for

R&S

®

ESU40 only

Firmware/software
Time Domain Scan

R&S

®

ESU-K53 1305.8509.02 yes

key

code

Triggered Scan for Coverage Meas.

R&S

®

ESPIK50 1106.4386.02 yes

key

code

8

With option R&S

®