Esu-b24 low-noise preamplifier – Atec Rohde-Schwarz-ESU-40 User Manual
Page 17
Version 02.00, October 2007
R&S
®
ESU EMI Test Receiver
17
R&S
®
ESU-B24 Low-Noise Preamplifier
R&S
®
ESU8
100 kHz to 8 GHz
R&S
®
ESU26
100 kHz to 26.5 GHz
Frequency range
R&S
®
ESU40
100 kHz to 40 GHz
Nominal gain
30
dB
Displayed average noise level (DANL) (analyzer mode)
RF attenuation = 0 dB, termination = 50 Ω, log. scaling, normalized to 1 Hz RBW,
preselection = OFF, preamplifier = OFF, low-noise preamplifier = OFF
f < 10 kHz: 10 Hz FFT Filter, trace average, sweep count = 20,
f ≥ 10 kHz: RBW = 1 kHz, VBW = 3 kHz, span = 0 Hz, sweep time = 50 ms, trace
average, sample detector, sweep count = 20, mean marker
all models
20 Hz
<–90 dBm
100 Hz
<–110 dBm
1 kHz
<–120 dBm
10 kHz
<–130 dBm
100 kHz
<–130 dBm
1 MHz
<–140 dBm
10 MHz
<–153 dBm
R&S
®
ESU8:
20 MHz ≤ f < 1 GHz
<–154 dBm, typ. –158 dBm
1 GHz ≤ f < 2 GHz
<–151 dBm, typ. –155 dBm
2 GHz ≤ f < 3 GHz
<–146 dBm, typ. –149 dBm
3 GHz ≤ f < 3.6 GHz
<–145 dBm, typ. –148 dBm
3.6 GHz ≤ f < 5 GHz
<–148 dBm, typ. –151 dBm
5 GHz ≤ f < 6 GHz
<–147 dBm, typ. –150 dBm
6 GHz ≤ f ≤ 8 GHz
<–144 dBm, typ. –147 dBm
R&S
®
ESU26:
20 MHz ≤ f < 1 GHz
<–152 dBm, typ. –156 dBm
1 GHz ≤ f < 2 GHz
<–150 dBm, typ. –154 dBm
2 GHz ≤ f < 3.6 GHz
<–145 dBm, typ. –148 dBm
3.6 GHz ≤ f < 10 GHz
<–147 dBm, typ. –150 dBm
10 GHz ≤ f < 18 GHz
<–145 dBm, typ. –149 dBm
18 GHz ≤ f < 22 GHz
<–142 dBm, typ. –145 dBm
22 GHz ≤ f ≤ 26.5 GHz
<–140 dBm, typ. –143 dBm
R&S
®
ESU40:
20 MHz ≤ f < 1 GHz
<–152 dBm, typ. –156 dBm
1 GHz ≤ f < 2 GHz
<–150 dBm, typ. –154 dBm
2 GHz ≤ f < 3.6 GHz
<–145 dBm, typ. –148 dBm
3.6 GHz ≤ f < 10 GHz
<–147 dBm, typ. –150 dBm
10 GHz ≤ f < 18 GHz
<–145 dBm, typ. –149 dBm
18 GHz ≤ f < 22 GHz
<–142 dBm, typ. –145 dBm
22 GHz ≤ f < 26.5 GHz
<–140 dBm, typ. –143 dBm
26.5 GHz ≤ f < 32 GHz
<–135 dBm, typ. –138 dBm
32 GHz ≤ f ≤ 40 GHz
<–133 dBm, typ. –135 dBm