General specifications – Atec Rohde-Schwarz-ESU-40 User Manual
Page 14
14 R&S
®
ESU EMI Test Receiver
Version 02.00, October 2007
General specifications
Display
21 cm LC TFT color display (8.4")
Resolution
800 × 600 pixels (SVGA resolution)
Pixel failure rate
<1 × 10
–5
Mass memory
hard
disk
Mass memory
option R&S
®
ESU-B18
hard disk replaced by a flash disk
Data storage
>500 instrument settings and traces
operating temperature range
+5° C to +40 °C
permissible temperature range
+0° C to +50 °C
Temperature
storage temperature range
–40°C to +70 °C
Climatic loading
+40 °C at 95 % relative humidity
(in line with EN 60068-2-30: 2000-02)
sinusoidal vibration
5 Hz to 150 Hz, max. 2 g at 55 Hz;
0.5 g from 55 Hz to 150 Hz;
in line with EN 60068-2-6: 1996-05,
EN 60068-2-30: 2000-02,
EN 61010-1, MIL-T-28800D, class 5
random vibration
10 Hz to 100 Hz, acceleration 1 g (RMS)
Mechanical resistance
shock
40 g shock spectrum, in line with
MIL-STD-810C and MIL-T-28800D,
classes 3 and 5
operation with external reference
2 years
Recommended calibration interval
operation with internal reference
1 year
RFI suppression
Complies with European EMC Directive
89/336/EEC and new EMC Directive
2004/108/EC including:
IEC/EN 61326 Class B (Emission)
CISPR 11/EN 55011/Group 1 Class B
(Emission)
IEC/EN 61326 Table A.1 (Immunity,
Industrial)
Power supply
AC supply
100 V to 240 V, 50 Hz to 400 Hz
Rated power consumption
3.1 A to 1.3 A or 500 VA
R&S
®
ESU8
typ. 130 VA
Power consumption
R&S
®
ESU26/40
typ. 150 VA
Safety
in line with EN 61010-1, UL 61010B-1,
CSA C22.2 No. 1010-1, IEC 61010-1
Test mark
VDE, GS, CSA, CSA-NRTL
Dimensions
W × H × D
435 mm × 192 mm × 460 mm
(17,13 in × 7,56 in × 18,11 in)
R&S
®
ESU8
15.6 kg (34.4 lb)
R&S
®
ESU26
16.7 kg (36.8 lb)
Weight net, w/o options, nominal
R&S
®
ESU40
17.0 kg (37.5 lb)