Atec Rohde-Schwarz-ESU-40 User Manual
Page 20
20 R&S
®
ESU EMI Test Receiver
Version 02.00, October 2007
Noise indication (receiver mode)
Nominal, calculated from DANL data
RF attenuation = 0 dB, termination = 50 Ω, average (AV) detector, preamplifier = OFF
all models
20 Hz, BW = 10 Hz
<27 dBµV
100 Hz, BW = 10 Hz
<7 dBµV
1 kHz, BW = 100 Hz
<7 dBµV
10 kHz, BW = 200 Hz
<0 dBµV
100 kHz, BW = 200 Hz
<0 dBµV
1 MHz, BW = 9 kHz
<7 dBµV
10 MHz, BW = 9 kHz
<–6 dBµV
R&S
®
ESU8:
20 MHz ≤ f < 30 MHz, BW = 9 kHz
<–8 dBµV
30 MHz ≤ f < 1 GHz, BW = 120 kHz
<3 dBµV
1 GHz ≤ f < 2 GHz, BW = 1 MHz
<12 dBµV
2 GHz ≤ f < 2.5 GHz, BW = 1 MHz
<16 dBµV
2.5 GHz ≤ f < 3 GHz, BW = 1 MHz
<19 dBµV
3 GHz ≤ f < 3.6 GHz, BW = 1 MHz
<25 dBµV
3.6 GHz ≤ f < 5 GHz, BW = 1 MHz
<19 dBµV
5 GHz ≤ f < 6 GHz, BW = 1 MHz
<20 dBµV
6 GHz ≤ f ≤ 8 GHz, BW = 1 MHz
<23 dBµV
R&S
®
ESU26:
20 MHz ≤ f < 30 MHz, BW = 9 kHz
<–5 dBµV
30 MHz ≤ f < 1 GHz, BW = 120 kHz
<6 dBµV
1 GHz ≤ f < 2 GHz, BW = 1 MHz
<15 dBµV
2 GHz ≤ f < 3 GHz, BW = 1 MHz
<20 dBµV
3 GHz ≤ f < 3.6 GHz, BW = 1 MHz
<25 dBµV
3.6 GHz ≤ f < 10 GHz, BW = 1 MHz
<20 dBµV
10 GHz ≤ f < 18 GHz, BW = 1 MHz
<22 dBµV
18 GHz ≤ f < 22 GHz, BW = 1 MHz
<25 dBµV
22 GHz ≤ f ≤ 26.5 GHz, BW = 1 MHz
<27 dBµV
R&S
®
ESU40:
20 MHz ≤ f < 30 MHz, BW = 9 kHz
<–5 dBµV
30 MHz ≤ f < 1 GHz, BW = 120 kHz
<6 dBµV
1 GHz ≤ f < 2 GHz, BW = 1 MHz
<15 dBµV
2 GHz ≤ f < 3 GHz, BW = 1 MHz
<20 dBµV
3 GHz ≤ f < 3.6 GHz, BW = 1 MHz
<25 dBµV
3.6 GHz ≤ f < 10 GHz, BW = 1 MHz
<20 dBµV
10 GHz ≤ f < 18 GHz, BW = 1 MHz
<22 dBµV
18 GHz ≤ f < 22 GHz, BW = 1 MHz
<25 dBµV
22 GHz ≤ f < 26.5 GHz, BW = 1 MHz
<27 dBµV
22 GHz ≤ f < 32 GHz, BW = 1 MHz
<32 dBµV
32 GHz ≤ f ≤ 40 GHz, BW = 1 MHz
<34 dBµV
RF attenuation = 0 dB, termination = 50 Ω, average (AV) detector, preamplifier = ON
5
all models
1 kHz, BW = 100 Hz
<–3 dBµV
10 kHz, BW = 200 Hz
<–10 dBµV
100 kHz, BW = 200 Hz
<–10 dBµV
1 MHz, BW = 9 kHz
<–3 dBµV
10 MHz, BW = 9 kHz
<–18 dBµV
R&S
®
ESU8:
20 MHz ≤ f < 30 MHz, BW = 9 kHz
<–18 dBµV
30 MHz ≤ f < 500 MHz, BW = 120 kHz
<–7 dBµV
500 MHz ≤ f < 1 GHz, BW = 120 kHz
<–4 dBµV
1 GHz ≤ f < 2 GHz, BW = 1 MHz
<5 dBµV
2 GHz ≤ f < 3 GHz, BW = 1 MHz
<8 dBµV
3 GHz ≤ f < 3.6 GHz, BW = 1 MHz
<13 dBµV
3.6 GHz ≤ f < 6 GHz, BW = 1 MHz
<5 dBµV
6 GHz ≤ f ≤ 8 GHz, BW = 1 MHz
<7 dBµV
5
With option R&S
®