8
Sub-harmonics
1
(At +10 dBm or maximum specified output
power, whichever is lower)
250 kHz to 10 GHz
None
> 10 GHz to 20 GHz
< –60 dBc
> 20 GHz to 44 GHz
< –45 dBc
Non-harmonics
2
(dBc at +10 dBm or maximum specified output
power, whichever is lower, for offsets > 3 kHz
[> 300 Hz with Option UNX])
Frequency
Spec
Typical
250 kHz to 250 MHz
–65
–72 for > 10 kHz offsets
> 250 MHz to 1 GHz
–80
–88
> 1 to 2 GHz
–74
–82
> 2 to 3.2 GHz
–68
–76
> 3.2 to 10 GHz
–62
–70
> 10 to 20 GHz
–56
–64
> 20 to 28.5 GHz
–52
–60
> 28.5 to 44 GHz
–48
–56
SSB phase noise (CW)
3
20 kHz offset from carrier (dBc/Hz)
Frequency
Spec
Typical
250 kHz to 250 MHz
–130
–134
> 250 to 500 MHz
–134
–138
> 500 MHz to 1 GHz
–130
–134
> 1 to 2 GHz
–124
–128
> 2 to 3.2 GHz
–120
–124
> 3.2 to 10 GHz
–110
–113
> 10 to 20 GHz
–104
–108
> 20 to 28.5 GHz
–100
–104
> 28.5 GHz
–96
–100
1. Sub-harmonics are defined as Carrier Freq / N). Specifications are typical for sub-harmonics beyond specified frequency range.
2. Performance is typical for spurs at frequencies above the maximum operating frequency of the instrument. Specifications
apply for CW mode, without modulation. In ramp sweep mode (Option 007), performance is typical for offsets > 1 MHz.
3. Measured at +10 dBm or maximum specified output power, whichever is less.