beautypg.com

Input/output descriptions, Front panel connectors – Atec Agilent-E8267D User Manual

Page 26

background image

26

RF output

Output impedance 50 Ω (nom)

Option 520

Precision APC-3.5 male or precision Type-N female

with Option 1ED

Options 532 and 544

Precision 2.4 mm male; plus 2.4(f) - 2.4(f) mm and

2.4(f) - 2.9(f) mm adaptors

ALC input

Used for negative external detector leveling

Nominal input impedance 120 kΩ, damage level ±15 V.

LF output

Outputs the internally generated LF source. Nominal output

impedance 50 Ω.

External input 1

Drives either AM, FM, or ΦM. Nominal input impedance 50

or 600 Ω, damage levels are 5 V

rms

and 10 V

peak

.

External input 2

Drives either AM, FM, or ΦM. Nominal input impedance 50

or 600 Ω, damage levels are 5 V

rms

and 10 V

peak

.

Pulse/trigger gate input Accepts input signal for external fast pulse modulation.

Also accepts external trigger pulse input for internal pulse

modulation. Nominal impedance 50 Ω. Damage levels are

5

V

rms

and 10 V

peak

.

Pulse video out

Outputs a signal that follows the RF output in all pulse modes.

TTL-level compatible, nominal source impedance 50 Ω.

Pulse sync out

Outputs a synchronizing pulse, nominally 50 ns width,

during internal and triggered pulse modulation. TTL-level

compatible, nominal source impedance 50 Ω.

Data clock input

Accepts a data clock signal to synchronize serial data for

use with internal baseband generator (Option 602).

Maximum rate 50 MHz. Damage levels are > +5.5 V and < –0.5 V.

Data input

Accepts serial data for use with internal baseband generator

(Option 602). Maximum rate 50 Mb/s. Data must be valid on

the falling edges of data clock (normal mode) or the symbol

sync (symbol mode). Damage levels are > +5.5 V and < –0.5 V.

I input

Accepts an "I" input either for I/Q modulation or for

wideband AM. Nominal input impedance 50 or 600 Ω.

Damage levels are 1 V

rms

and 5 V

peak

.

Q input

Accepts a "Q" input for I/Q modulation. Nominal input

impedance 50 or 600 Ω. Damage levels are 1 V

rms

and 5 V

peak

.

Symbol sync input

Accepts symbol sync signal for use with internal baseband

generator (Option 602). Symbol sync might occur once per

symbol or be a single, one bit wide pulse to synchronize

the first bit of the first symbol. Maximum rate 50 MHz.

Damage levels are > +5.5 V and < –0.5 V.

Input/Output Descriptions

Front panel connectors

(All connectors are BNC female unless
otherwise noted.)

1

1. Digital inputs and outputs are 3.3 V CMOS unless indicated otherwise. Inputs will accept 5 V CMOS, 3 V CMOS or TTL voltage levels.