Input/output descriptions, Front panel connectors – Atec Agilent-E8267D User Manual
Page 26
![background image](/manuals/318647/26/background.png)
26
RF output
Output impedance 50 Ω (nom)
Option 520
Precision APC-3.5 male or precision Type-N female
with Option 1ED
Options 532 and 544
Precision 2.4 mm male; plus 2.4(f) - 2.4(f) mm and
2.4(f) - 2.9(f) mm adaptors
ALC input
Used for negative external detector leveling
Nominal input impedance 120 kΩ, damage level ±15 V.
LF output
Outputs the internally generated LF source. Nominal output
impedance 50 Ω.
External input 1
Drives either AM, FM, or ΦM. Nominal input impedance 50
or 600 Ω, damage levels are 5 V
rms
and 10 V
peak
.
External input 2
Drives either AM, FM, or ΦM. Nominal input impedance 50
or 600 Ω, damage levels are 5 V
rms
and 10 V
peak
.
Pulse/trigger gate input Accepts input signal for external fast pulse modulation.
Also accepts external trigger pulse input for internal pulse
modulation. Nominal impedance 50 Ω. Damage levels are
5
V
rms
and 10 V
peak
.
Pulse video out
Outputs a signal that follows the RF output in all pulse modes.
TTL-level compatible, nominal source impedance 50 Ω.
Pulse sync out
Outputs a synchronizing pulse, nominally 50 ns width,
during internal and triggered pulse modulation. TTL-level
compatible, nominal source impedance 50 Ω.
Data clock input
Accepts a data clock signal to synchronize serial data for
use with internal baseband generator (Option 602).
Maximum rate 50 MHz. Damage levels are > +5.5 V and < –0.5 V.
Data input
Accepts serial data for use with internal baseband generator
(Option 602). Maximum rate 50 Mb/s. Data must be valid on
the falling edges of data clock (normal mode) or the symbol
sync (symbol mode). Damage levels are > +5.5 V and < –0.5 V.
I input
Accepts an "I" input either for I/Q modulation or for
wideband AM. Nominal input impedance 50 or 600 Ω.
Damage levels are 1 V
rms
and 5 V
peak
.
Q input
Accepts a "Q" input for I/Q modulation. Nominal input
impedance 50 or 600 Ω. Damage levels are 1 V
rms
and 5 V
peak
.
Symbol sync input
Accepts symbol sync signal for use with internal baseband
generator (Option 602). Symbol sync might occur once per
symbol or be a single, one bit wide pulse to synchronize
the first bit of the first symbol. Maximum rate 50 MHz.
Damage levels are > +5.5 V and < –0.5 V.
Input/Output Descriptions
Front panel connectors
(All connectors are BNC female unless
otherwise noted.)
1