beautypg.com

Atec Agilent-E8267D User Manual

Page 6

background image

1. Specifications apply in CW and list/step sweep modes over the 15 to 35 ºC temperature range, with attenuator hold off
(normal operating mode). Degradation outside this range, for ALC power levels > –5 dBm, is typically < 0.3 dB. In ramp sweep
mode (with Option 007), specifications are typical. For instruments with Type-N connectors (Option 1ED), specifications apply
to 18 GHz only. From 18 to 20 GHz, typical level accuracy degrades by 0.2 dB. Specifications do not apply above the
maximum specified power.
2. If external inputs are used, specification applies with input level

(I

2

+ Q

2

) = 0.3 V

rms

and I/Q modulator attenuation is

internally optimized based on input levels.
3. Measured with symbol rate > 10 kHz and power ≤ 0 dBm.
4. Relative to ALC on, after power search is executed. When applying external I/Q signals with ALC off, output level will vary
directly with I/Q input level.
5. Options 532 and 544: 0.02 dB/°C (typ) above 2 GHz.
6. Compatible with Agilent EPM Series (E4418B and E4419B) power meters.

6

Level accuracy

1

(dB)

Frequency

> +10 dBm

+10 to –10 dBm

< –10 to –70 dBm < –70 to –90 dBm

250 kHz to 2 GHz

±0.6

±0.6

±0.7

±0.8

> 2 to 20 GHz

±0.8

±0.8

±0.9

±1.0

>20 to 32 GHz

±1.0

±0.9

±1.0

±1.7

> 32 to 44 GHz

±1.0

±0.9

±1.5

±2.0

CW level accuracy with I/Q modulation (With PRBS modulated data) (relative to CW)

2

With ALC on:
QAM or QPSK formats

3

± 0.2 dB

Constant-amplitude formats (FSK, GMSK, etc)

± 0.2 dB

With ALC off:

4

± 0.2 dB (typ)

Level accuracy (measured)

Resolution

0.01 dB

Temperature stability

0.01 dB/ °C (typ)

5

User flatness correction
Number of points

2 to 1601 points/table

Number of tables

Up to 10,000, memory limited

Path loss

Arbitrary, within attenuator range

Entry modes

Remote power meter

6

, remote bus, manual

(user

edit/view)

Output impedance

50 Ω (nom)

SWR (internally leveled)
250 kHz to 2 GHz

< 1.4:1 (typ)

> 2 GHz to 20 GHz

< 1.6:1 (typ)

> 20 GHz

< 1.8:1 (typ)

Leveling modes

Internal leveling, external detector leveling,

millimeter source module, ALC off

0 5 10 15 20

Frequency (GHz)

Option 520 level accuracy at -110 dBm (measured)

Error (dB)

0.4
0.3
0.2
0.1

0

-0.1
-0.3
-0.4
-0.5

0 4 8 12 16 20 24 28 32 36 40 44

Frequency (GHz)

Option 544 level accuracy at -110 dBm (measured)

Error (dB)

0.4
0.3
0.2
0.1

0

-0.1
-0.3
-0.4
-0.5