Output

Power

1,2

(dBm)

Frequency range

spec (typ)

Option 520
250 kHz to 3.2 GHz

–130 to +13 (+16)

250 kHz to 3.2 GHz with Option UNW

–130 to +9 (+13)

250 kHz to 3.2 GHz with Option 1EH

–130 to +10 (+13)

3

250 kHz to 3.2 GHz with Options UNW and 1EH

–130 to +7

(+12)

3

> 3.2 to 10 GHz

–130 to +18 (+23)

4

> 10 to 20 GHz

–130 to +18 (+22)

4

Options 532 and 544
250 kHz to 3.2 GHz

–130 to +12 (+15)

250 kHz to 3.2 GHz with Option UNW

–130 to +8 (+12)

250 kHz to 3.2 GHz with Option 1EH

–130 to +9

(+12)

3

250 kHz to 3.2 GHz with Options UNW and 1EH

–130 to +6

(+11)

3

> 3.2 to 10 GHz

–130 to +14

(+21)

4

> 10 to 20 GHz

–130 to +14

(+18)

4

> 20 to 32 GHz

–130 to +14

(+18)

5

> 32 to 40 GHz

–130 to +12 (+18)

5

> 40 to 44 GHz

–130 to +10 (+13)

5

Step attenuator

6

0 to 115 dB in 5 dB steps

Maximum available power in CW mode (measured)

Attenuator hold range
Minimum

From –15 dBm to maximum specified

output power with step attenuator in

0 dB position. Can be offset using step

attenuator.

Amplitude switching speed

7

ALC on or off (without power search)

< 3 ms (typ)

1. Maximum power specification is warranted from 15 to 35 ºC, and is typical from 0 to 15 ºC. Maximum power over the
35 to 55 ºC range typically degrades less than 2 dB unlessotherwise stated.
2. With I/Q modulation on, maximum power specification is typical. With external inputs enabled,

(I

2

+ Q

2

) > 0.2 V

rms

.

3. With harmonic filters switched off. With filters on, maximum output power is reduced 3 dB for frequencies below 2 GHz.
4. With I/Q modulation on, maximum power specification is typically reduced 3 dB.
5. Maximum power over the 35 to 55 ºC range typically degrades less than 4 dB. With I/Q modulation on, maximum power
specification is typically reduced 5 dB.
6. The step attenuator provides coarse power attenuation to achieve low power levels. Fine power level adjustment is
provided by the ALC (Automatic Level Control) within the attenuator hold range.
7. To within 0.1 dB of final amplitude within one attenuator range. Add 10 to 50 ms when using power search.

Option 520 output power (measured)

Frequency (MHz)

26

24

22

20

18

16

14

0 5000 10000 15000 20000

Power (dBm)

0 4 8 12 16 20 24 28 32 36 40 44

Frequency (GHz)

Option 544 output power (measured)

Power (dBm)

26

24

22

20

18

16

14