Clock Reference
External reference output
81150A
81160A
Frequency
Accuracy
Stability
Aging
10 MHz typ.
± 50 ppm
± 2 ppm, 0 to 50 ºC
± 1 ppm per year
10 MHz spec.
± 10 ppm
± 2 ppm, 0 to 55 °C
± 1 ppm per year
Output level
1 V nom.
1 V nom.
Impedance
50 Ω nom., AC coupled
50 Ω nom., AC coupled
Connector
BNC, rear panel
BNC, rear panel
External reference input
81150A
81160A
Lock range
10 MHz ± 500 ppm
10 MHz ± 500 ppm typ.
Input level
200 mV
pp
to 5 V
pp
200 mV
PP
to 5 V
PP
typ.
Impedance
1 kΩ nom., AC coupled
1 kΩ nom., AC coupled
Connector
BNC, rear panel
BNC, rear panel
Internal frequency characteristics
81150A
81160A
Accuracy
± 50 ppm
±10 ppm
Stabilty
± 2 ppm, 0 to 50 ºC
± 2 ppm, 0 to 55 ºC
Aging
± 1 ppm per year
± 1 ppm per year
External Input
A common external input is provided for both channels on the front panel. The external
input is used for external trigger or external gate modes.
81150A
81160A
Frequency range
DC to 120 MHz
DC to 330 MHz
Input range
-10 V to +10 V
- 5 V to + 5 V
Maximum input amplitude
10 V
PP
10 V
PP
Input sensitivity
200 mV
PP
Hysteresis low: 200 mV
PP
Hysteresis high: 350 mV
PP
Threshold
Range
Resolution
-10 V to 10 V
100 mV
- 5 V to + 5 V
100 mV
Impedance
Selectable 10 kΩ /50 Ω, DC coupled
Selectable, 1 kΩ/ 50 Ω, DC coupled
Slope
Selectable, rising/falling/both
Selectable, rising/falling/both
Pulse width
> 3.3 ns
> 1.3 ns
Transition time
< 100 ns
Connector
BNC, front panel
BNC, front panel
Frequency counter
The frequency applied to external input is measured
Gate time, up to 330 MHz
—
1 s, fi x
Accuracy
—
See clock reference specifi cations