beautypg.com

Atec Agilent-81150A-81160A User Manual

Page 2

background image

2

• 1 µHz – 120 MHz pulse generation with variable rise/fall time

• 1 µHz – 240 MHz sine waveform output

• 14-bit, 2 GSa/s arbitrary waveforms

• 512k samples deep arbitrary waveform memory per channel

• Pulse, sine, square, ramp, noise and arbitrary waveforms

• Noise, with selectable crest factor, and signal repetition time of 26 days

• FM, AM, PM, PWM, FSK modulation capabilities

• 1 or 2 channel, coupled and uncoupled

• Differential

outputs

• Two selectable output amplifi ers:

◦ High bandwidth amplifier

Amplitude: 50 mV

PP

to 5 V

PP

; 50 Ω into 50 Ω

100 mVpp to 10 V

PP

; 50 Ω into open

Voltage window: ± 5 V; 50 Ω into 50 Ω

± 10 V; 50 Ω into open

± 9 V; 5 Ω into 50 Ω

◦ High voltage amplifier

Amplitude: 100 mV

PP

to 10 V

PP

; 50 Ω into 50 Ω,

200

mV

PP

to 20 V

PP

; 5 Ω into 50 Ω, or 50 Ω into open

Voltage window: ± 10 V; 50 Ω into 50 Ω

± 20 V; 5 Ω into 50 Ω or 50 Ω into open

• Glitch free change of timing parameters

(delay, frequency, transition time, width, duty cycle)

• Programming language compatible with Agilent 81101A, 81104A, 81105A,

81110A, 81130A and 81160A

• ISO 17025 and Z540.3 calibration

• LXI class C (rev. 1.1) compliant

• Optional pattern generator:

◦ Ideal and arbitrary bit shaped pattern up to 120 Mbit/s

◦ Two, three or four level signals

◦ PRBS up to 2

31

◦ 16 Mbit pattern memory

◦ Pass through pattern for combined and physical and protocol test up to 10 Mbit/s

The 81150A Pulse Function
Arbitrary Noise Generator
at a Glance

811 A

50