10 measurement time – Atec Anritsu-ME7838A User Manual
Page 8
8
PN: 11410-00593 Rev. D
ME7838A BB/mm-Wave VNA TDS
Broadband Specifications and Performance
VectorStar ME7838A BB/mm-Wave VNA
2.10 Measurement Time
Measurement times include sweep time, retrace time, and band-switching time. Typical.
Measurement Time (ms)
Full Band, 70 kHz to 110 GHz, Display ON, and ALC ON.
Calibration
IFBW
Measurement Time (ms)
a
a. Measurement times are for ME7838A Broadband and ME7838A Millimeter-Wave Systems.
1-port calibration
1 MHz
80
100
350
700
30 kHz
90
160
600
1500
10 kHz
110
240
1100
2600
1 kHz
470
1600
10,000
25,000
10 Hz
47,000
160,000
1,000,000
2,500,000
2-port calibration
1 MHz
160
200
700
1400
30 kHz
180
320
1200
3000
10 kHz
220
480
2200
5200
1 kHz
940
3200
20,000
50,000
10 Hz
94,000
320,000
2,000,000
5,000,000
Measurement Time (ms) vs. System Dynamic Range (dB)
Full Band, Display ON, and ALC ON.
Calibration
401 Points
Measurement Time
Achieved System Dynamic
Range (Opt 062 at 54 GHz)
IFBW and Averaging Used
Uncorrected or
1-port calibration
110
470
77
87
10 kHz/no avg
1 kHz/no avg
2-port calibration
220
940
77
87
10 kHz/no avg
1 kHz/no avg
401 Points
1,601 Points
10,001 Points
25,000 Points