beautypg.com

13 measurement examples – Atec Anritsu-ME7838A User Manual

Page 10

background image

10

PN: 11410-00593 Rev. D

ME7838A BB/mm-Wave VNA TDS

Broadband Specifications and Performance

VectorStar ME7838A BB/mm-Wave VNA

2.13 Measurement Examples

The following figures are measurement examples of the ME7838A Broadband system performance.

10 GHz

110 GHz

Dynamic range of ME7838A system at the W1 1 mm coaxial test port from 70 kHz to 125 GHz.

Pow er Meter vs a1/1 Receiver

–10.5

–10.4

–10.3

–10.2

–10.1

–10

–9.9

–9.8

–9.7

–9.6

–9.5

0

10

20

29

39

48

58

67

76

86

95

105

114

125

Frequency (GHz)

dB

m

|a1/1|
Power Meter

An example of power measurement agreement: power sensor vs. ME7838A a1 reference receiver.

Power sweep range at 77 GHz. By using detection and power control

inside the 3743A millimeter-wave module; improved accuracy,

linearity and range can be achieved.

Power sweep range at 94 GHz demonstrating

greater than 50 dB of control.