2. Specifications for Broadband Configuration
Broadband Specifications and Performance
2.1 ME7838A Broadband System Configuration
ME7838A Broadband Hardware Configuration
The ME7838A broadband VNA system provides single sweep coverage from 70 kHz to 110 GHz and is operational from 40 kHz to 125 GHz. It
consists of the following items:
VNA
MS4647A VectorStar VNA, 70 kHz to 70 GHz with Option 007, Option 070, and Option 080/081
Test Set
3739B Broadband Test Set and interface cables
mm-Wave Modules
3743A Millimeter-Wave Modules, 2 each
ME7838A Broadband System Major Options
The major ME7838A broadband VNA system options are:
Option 002
MS4640A-002, Time Domain
Option 041
MS4640A-041, Noise Figure
Option 051
MS4647A-051, External VNA Direct Access Loops
Option 061
MS4647A-061, Active Measurement Suite, with 2 Attenuators
Option 062
MS4647A-062, Active Measurement Suite, with 4 Attenuators
Bias Tees
SC8215 and SC7287 Kelvin Bias Tees
2.2 System and Receiver Dynamic Range, Noise Floor
(Excludes localized spurious responses and crosstalk)
System Dynamic Range
System dynamic range is measured as the difference between maximum port power and the RMS noise
floor in a 10 Hz bandwidth and no averaging (ports terminated).
Noise Floor
Noise floor is calculated as the difference between maximum rated port power and system dynamic range.
Receiver Dynamic Range
Receiver Dynamic Range is calculated as the difference between the receiver compression level and the
noise floor at Ports 1 or 2.
Normalizing Measurement
Normalizing measurement made with a through line connection, with its effects compensated for. The
cables between the VNA and the 3743A modules are assumed to be the 806-206 1.85 mm cable (61 cm,
24 in long) or the 806-209 1.85 mm cable (91 cm, 36 in long). All figures are typical.
Frequency Range
System Dynamic Range (dB)
a
a. Excludes localized spurious responses and crosstalk.
70 to 300 kHz
93
90
89
86
–83
–80
0.3 to 2 MHz
103
100
103
102
–93
–90
2 to 10 MHz
115
112
115
114
–105
–102
0.01 to 2.5 GHz
120
116
121
122
–110
–109
2.5 to 24 GHz
110
105
121
121
–110
–108
24 to 54 GHz
110
107
124
123
–114
–113
54 to 60 GHz
108
108
122
122
–112
–112
60 to 67 GHz
108
108
117
117
–107
–107
67 to 80 GHz
108
108
120
120
–110
–110
80 to 85 GHz
107
107
123
123
–113
–113
85 to 90 GHz
107
107
121
121
–111
–111
90 to 95 GHz
109
109
121
121
–111
–111
95 to 105 GHz
107
107
117
117
–107
–107
105 to 110 GHz
109
109
122
122
–112
–112
110 to 120 GHz
b
b. 110 to 125 GHz frequency range is available as operational.
107
107
115
115
–110
–110
120 to 125 GHz
104
104
112
112
–107
–107
Receiver Dynamic Range (dB)
a
Noise Floor (dBm)
a
ME7838A
ME7838A
Option 062
ME7838A
ME7838A
Option 062
ME7838A
ME7838A