beautypg.com

3 test port power, receiver compression – Atec Anritsu-ME7838A User Manual

Page 4

background image

4

PN: 11410-00593 Rev. D

ME7838A BB/mm-Wave VNA TDS

Broadband Specifications and Performance

VectorStar ME7838A BB/mm-Wave VNA

2.3 Test Port Power, Receiver Compression

Port power control is provided by the base VNA for frequencies below 54 GHz, and by the 3743A mm-Wave module for frequencies greater than

54 GHz. Receiver compression point is defined as the port power level beyond which the response may be compressed more than 0.2 dB relative

to normalization level. 10 Hz IF bandwidth used to remove trace noise effects. All typical.

Frequency Range

Port Power

Receiver Compression

a

a. Using the 806-206 1.85 mm (61 cm, 24 in long) test port cables or the 806-209 1.85 mm (91 cm, 36 in long) test port cables between the

VNA and the 3743A mm-Wave modules.

70 to 300 kHz

10

10

6

6

0.3 to 2 MHz

10

10

10

12

2 to 10 MHz

10

10

10

12

0.01 to 2.5 GHz

10

7

11

13

2.5 to 24 GHz

0

–3

11

13

24 to 54 GHz

–4

–6

10

10

54 to 60 GHz

–4

–4

10

10

60 to 67 GHz

1

1

10

10

67 to 80 GHz

–2

–2

10

10

80 to 85 GHz

–6

–6

10

10

85 to 90 GHz

–4

–4

10

10

90 to 95 GHz

–2

–2

10

10

95 to 105 GHz

0

0

10

10

105 to 110 GHz

–3

–3

10

10

110 to 120 GHz

b

b. 110 to 125 GHz frequency range is available as operational.

–3

–3

5

5

120 to 125 GHz

b

–3

–3

5

5

Power Range, Accuracy, Linearity, and Resolution

Accuracy is defined at –10 dBm or max rated power, whichever is lower. Linearity is defined as the incremental error between the accuracy test

power level and 5 dB below. Typical.

Frequency

Range (dB)

Accuracy

(dB)

Linearity

(dB)

Resolution

(dB)

ME7838A

ME7838A Option 062

70 to 300 kHz

-25 to +10

-85 to +10

±1.5

±1.5

0.01

0.3 to 2 MHz

-25 to +10

-85 to +10

±1.5

±1.5

0.01

2 to 10 MHz

-25 to +10

-85 to +10

±1.5

±1.5

0.01

.01 to 2.5 GHz

-25 to +10

-85 to +8

±1.0

±1.0

0.01

2.5 to 24 GHz

-25 to 0

-85 to -3

±1.0

±1.0

0.01

24 to 54 GHz

-30 to -4

-90 to -6

±1.5

±1.0

0.01

54 to 60 GHz

-55 to -4

-55 to -4

±2.0

±1.5

0.01

60 to 67 GHz

-55 to +1

-55 to +1

±2.0

±1.5

0.01

67 to 80 GHz

-55 to -2

-55 to -2

±2.0

±1.5

0.01

80 to 85 GHz

-55 to -6

-55 to -6

±2.0

±1.5

0.01

85 to 90 GHz

-55 to -4

-55 to -4

±2.0

±1.5

0.01

90 to 95 GHz

-55 to -2

-55 to -2

±2.0

±1.5

0.01

95 to 105 GHz

-55 to 0

-55 to 0

±3.0

±2.0

0.01

105 to 110 GHz

-50 to -3

-55 to -3

±3.0

±2.0

0.01

110 to 120 GHz

a

a. 110 to 125 GHz frequency range is available as operational.

-40 to -3

-40 to -3

±4.0

±3.0

0.01

120 to 125 GHz

a

-40 to -3

-40 to -3

±4.0

±3.0

0.01

Max Power

ME7838A

Max Power

ME7838A Option 062

Compression

ME7838A

Compression

ME7838A Option 062