Definitions – Atec Anritsu-ME7838A User Manual
Page 2
2
PN: 11410-00593 Rev. D
ME7838A BB/mm-Wave VNA TDS
Table of Contents
VectorStar ME7838A BB/mm-Wave VNA
Description
Page
1. Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2. Specifications for Broadband Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3. Specifications for Waveguide Band Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4. VectorStar ME7838A Waveguide Bands from 50 GHz to 750 GHz (1.1 THz). . . . . . . . . . . . . . . . 21
5. Standard Capabilities for All Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6. Mechanical and Environmental. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7. Calibration and Correction Capabilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
8. Mechanical Calibration/Verification Kits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
9. Test Port Cables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
10. Precision Adapters, Attenuators, and Other Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
11. Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
12. Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
1. Definitions
All specifications and characteristics apply under the following conditions, unless otherwise stated:
Warm-Up Time
After 90 minutes of warm-up time, where the instrument is left in the ON state.
Temperature Range
Over the 25 °C ± 5 °C temperature range.
Error-Corrected Specifications
For error-corrected specifications, over 23 °C ± 3 °C, with < 1 °C variation from calibration temperature.
For error-corrected specifications are warranted and include guard bands, unless otherwise stated.
Typical Performance
"Typical" specifications describe expected, but not warranted, performance based on sample testing.
Typical performance indicates the measured performance of an average unit and do not guarantee the
performance of any individual product. "Typical" specifications do not account for measurement uncertainty
and are shown in parenthesis, such as (-102 dB), or noted as Typical.
User Cables
Specifications do not include effects of any user cables attached to the instrument.
Discrete Spurious Responses
Specifications may exclude discrete spurious responses.
Internal Reference Signal
All specifications apply with internal 10 MHz Crystal Oscillator Reference Signal.
Characteristic Performance
Characteristic performance indicates a performance designed-in and verified during the design phase. It
does include guard-bands and is not covered by the product warranty.
Below 300 kHz
All uncertainties below 300 kHz are typical.
Recommended Calibration Cycle
12 months
Interpolation Mode
All specifications are with Interpolation Mode Off.
Specifications Subject to Change
All specifications subject to change without notice. For the most current data sheet, please visit the Anritsu
web si
.