beautypg.com

Standard capabilities for all configurations, Mechanical and environmental – Atec Anritsu-ME7838A User Manual

Page 25

background image

VectorStar ME7838A BB/mm-Wave VNA

Standard Capabilities

ME7838A BB/mm-Wave VNA TDS

PN: 11410-00593 Rev. D

25

5. Standard Capabilities for All Configurations

Standard Capabilities

For standard capabilities of the VectorStar VNAs, please see the VectorStar MS4640A Series VNA Technical Data Sheet and Configuration

Guide – 11410-00432, available at www.anritsu.com.

6. Mechanical and Environmental

MS4640A Vector Network Analyzer

Dimensions without rack mount option.

Height

267 mm body (6u)

286 mm between feet outer edges

Width

426 mm body

457 mm between feet outer edges

487 mm between front panel handles outer edges

Depth

502 mm body

591 mm between handle and foot outer edges

Weight

< 28 kg (< 60 lbs), Typical weight for a fully-loaded MS4647A VNA

3739B Broadband/Millimeter-Wave Test Set

Dimensions without rack mount option.

Height

89 mm body (2u)

108 mm between feet outer edges

Width

426 mm body

457 mm between feet outer edges

487 mm between front panel handles outer edges

Depth

502 mm body

591 mm between handle and foot outer edges

Weight

5.75 kg

3743A Millimeter-Wave Module

Height

21.5 mm

Width

54 mm

Depth

55.3 mm

Weight

0.27 kg

Environmental – Operating

Conforms to MIL-PRF-28800F (Class 3)

Temperature Range

0 ºC to +50 ºC without error codes*

* Except for ‘unleveled’ error messages that may occur at the extreme edges of the temperature range

above.

Relative Humidity

5 % to 95 % at +40 ºC, Non-condensing

Altitude

4,600 m (15,000 feet)

Environmental – Non-Operating

Temperature Range

–40 ºC to +75 ºC

Relative Humidity

0 % to 90 % at +65 ºC , Non-condensing

Altitude

15,200 m (49,000 feet)

Relative Humidity

0 % to 90 % at +65 °C, Non-condensing

EMI

Meets the emissions and immunity requirements of:

EN55011/2007 Class A, Group 1

EN61000-4-2: 1998 (± 4 kV CD, 8k AD)

EN61000-4-3: 2008 (80 MHz to 2700 MHz @ 3 V/m)

EN61000-4-4: 2004 (500V SL, 1000V PL)

EN61000-4-5: 2006 (2 kV L-E, 1 kV L-L)

EN61000-4-6:2007 (0.15 MHz to 80 MHz, 3 V)

EN61000-4-11:2004 (1 cycle, 100 %)