beautypg.com

Hp 8753e specifications – Atec Agilent-8753E User Manual

Page 8

background image

8

HP 8753E specifications

1.

At 25˚ C ±5˚ C, relative to 0 dBm output power for the HP 8753E, +10 dBm output power
for the HP 8753E Option 011.

2.

Typical below 300 kHz.

3.

16 MHz to 3 GHz.

4.

16 MHz to 2 GHz.

5. Typical from 2 to 3 GHz for instruments with Option 075.
6.

+8 dBm with Option 075.

7.

Test performed on port 1 only.

Test-port output characteristics

7

Frequency characteristics

Range

30 kHz to 3 GHz (6 GHz with Opt. 006)

Resolution

1 Hz

Stability

typically ±7.5 ppm 0° to 55° C

typically ±3 ppm/year

With Option 1D5

typically ±0.05 ppm 0° to 55° C

typically ±0.5 ppm/year

Accuracy

±10 ppm at 25° C ±5° C

Power range

–85 to +10 dBm

2,6

Resolution

0.05 dB

Level accuracy

1,2,5

±1.0 dB

Level linearity

1,2,5

(–15 dBm to +5 dBm) ±0.2 dB

(5 dBm to 10dBm)

6

±0.5 dB

Impedance

50

Ω; typically

≥16 dB RL (<1.38 SWR) to 3 GHz
≥14 dB RL (<1.50 SWR) to 6 GHz

Spectral purity

2nd harmonic

3

<–25 dBc at 10

6

dBm

<–40 dBc at 0 dBm (typical)

<–50 dBc at –10 dBm (typical)

3rd harmonic

4

<–25 dBc at 10

6

dBm

<–40 dBc at 0 dBm (typical)

<–50 dBc at –10 dBm (typical)

Nonharmonic spurious

Mixer related

<–30 dBc at 10

6

dBm (typical)

<–55 dBc at –10 dBm (typical)

Test port input characteristics

Frequency range

30 kHz to 3 GHz (6 GHz with Opt. 006)

Average noise level

2

–82 dBm (3 kHz BW, <3 GHz)

–102 dBm (10 Hz BW, <3 GHz)

–110 dBm (10 Hz BW, <3 GHz) (typical)

–77 dBm (3 kHz BW, 3 to 6 GHz)

–97 dBm (10 Hz BW, 3 to 6 GHz)

–105 dBm (10 Hz BW, 3 to 6 GHz) (typical)

Maximum input level

10 dBm

Damage level

26 dBm or 35 VDC

Impedance, 50 ohms

>10 dB RL, 30 kHz to 50 kHz

2

>20 dB RL, 50 kHz to 300 kHz

2

>18 dB RL, 300 kHz to 1.3 GHz

>16 dB RL, 1.3 GHz to 3 GHz

>14 dB RL, 3 GHz to 6 GHz

Frequency response

2,5

± 1.0 dB, 300 kHz to 3 GHz

(25° + 5° C)

± 2.0 dB, 3 GHz to 6 GHz

Harmonics (Option 002)

2nd harmonic

3

<–15 dBc at +8 dBm

<–35 dBc at 0 dBm (typical)

<–45 dBc at –15 dBm (typical)

3rd harmonic

4

<–30 dBc at +8 dBm

<–50 dBc at 0 dBm (typical)

<–50 dBc at –15 dBm (typical)

Harmonic measurement accuracy

(25° ±5° C)

16 MHz to 3 GHz

± 1.5 dB

3 GHz to 6 GHz

± 3 dB (with Opt. 006)

Harmonic measurement dynamic range

–40 dBc (output = –10 dBm,

input = <–15 dBm)

A/D

Frequency

Reference

Synthesizer/

Pulse Gen.

Main

Processor

ROM/RAM

I/O

Front

Panel

Display

HP-IB

Fast

Processor

Phase Lock

Receiver

3.8 to

6.8 GHz

3.8 GHz

Source

4 kHz

RF Out

30 kHz

to 3 GHz

B

A

R

R out

R in

1 MHz

HP 8753E block diagram