beautypg.com

Hp 8753e supplemental characteristics, cont’d, General characteristics – Atec Agilent-8753E User Manual

Page 17

background image

17

HP 8753E supplemental characteristics, cont’d

Transfer formats
Binary (internal 48-bit floating point complex format)
ASCII 32- or 64-bit IEEE 754 floating point format

User-accessible graphics
Using a subset of HP graphics language (HP-GL),
vector or text graphics may be written on the HP 8753E via
HP-IB. Up to 5 kbytes of data can be stored at one time
(4 bytes per vector, 2 bytes per character).

Interface function codes
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT0, C1,
C2, C3, C10, E2

General characteristics

Front panel connectors

HP 8753E test ports (without Option 011)

Connector type

7 mm, precision

Impedance

50 ohms (nominal)

Connector conductor depth

0.000 to 0.003 in.

Option 011 test ports

Connector type

Type-N

Impedance

50 ohms (nominal)

Connector center pin protrusion

0.204 to 0.207 in.

Option 075 test ports

Connector type

Type-N

Impedance

75 ohms (nominal)

Connector center pin protrusion

0.204 to 0.207 in.

Probe power

+15V ±2% 400 mA (combined

load for both probe connections)

–12.6V ±5.5% 300 mA (combined

load for both probe connections)

Rear panel connectors

External reference frequency input (EXT REF INPUT)

Frequency

1, 2, 5, and 10 MHz (±200 Hz at

10 MHz)

Level

–10 dBm to +20 dBm, typical

Impedance

50 ohms

Connector

BNC (f)

High-stability frequency reference output (Option 1D5)

Frequency

10.0000 MHz

Frequency stability

±0.05 ppm

(0° C to 55° C)

Daily aging rate

(after 30 days)

<3x10

–9

/day

Yearly aging rate

0.5 ppm/year

Output

0 dBm minimum

Nominal output impedance

50

Connector

BNC (f)

External auxiliary input (AUX INPUT)

Input voltage limits

–10V to +10V

External AM input (EXT AM)

±1 volt into a 5 k

Ω resistor,

1 kHz maximum, resulting in

8 dB/volt amplitude modulation.

BNC (f) connector.

External trigger(EXT TRIGGER)

Triggers on a negative TTL

transition or contact closure

to ground. BNC (f) connector.

Test sequence output (TEST SEQ)
By default, this connector outputs a TTL end-of-sweep
signal. It can also be programmed by the user in a test
sequence to output a user-defined TTL signal. BNC (f)
connector.

Limit test output (LIMIT TEST)
This connector outputs a TTL signal of the limit test results.
Pass: TTL high. Fail: TTL low. BNC (f)
connector.

Test-port bias input (BIAS CONNECT)

Maximum voltage

+30 VDC

Maximum current

(no degradation in RF specs

)

±200 mA

Maximum current

±1 A

Connector

BNC (f)

VGA video output (EXT MON)
This connector drives external VGA monitors.

HP-IB
This connector allows communications with compatible
devices including external controllers, printers, plotters,
disk drives, and power meters.

Parallel port
This 25-pin female connector is used with parallel (or
Centronics interface) peripherals such as printers and
plotters. It can also be used as a general purpose I/O port,
with control provided by test sequencing functions.

RS-232C
This 9-pin male connector is used with serial peripherals
such as printers and plotters.

DIN keyboard
This connector is used for adding an IBM PC-AT
compatible keyboard for titles and remote front-panel
operation.

Test set interconnect
This connector is used to connect an HP 8753E
Option 011 to the HP 85046A/B or 85047A test set.
On other HP 8753E analyzers, you can use signal
levels on this connector for sequencing or general
purpose I/O applications.