Atec Agilent-8753E User Manual
Page 4
![background image](/manuals/318371/4/background.png)
4
System performance summary
HP 8753E (50-ohm systems) type-N test ports
1.
90 dB, 30 kHz to 50 kHz.
2.
100 dB, 300 kHz to 16 MHz due to fixed spurs.
3.
These measurement uncertainty curves utilize an RSS model for the contribution of random
errors such as noise, typical connector repeatabilities, and test set switch; with a worst-case
model for the contributions of dynamic accuracy and residual systematic errors.
4.
The graphs shown for transmission measurements assume a well-matched device (S11 = S22 = 0).
device with more than 6 dB insertion loss.
5.
Typical below 300 kHz.
The following specifications describe the system perfor-
mance of the HP 8753E network analyzer with an
integrated 50-ohm S-parameter test set configuration.
System hardware includes the following:
Network analyzer
HP 8753E Option 006
Calibration kit
HP 85032B
Test-port cables
HP 11857D
Dynamic range
These specifications apply to transmission measurements
in the 30 kHz to 6 GHz frequency range at 10 Hz IF BW
with full two-port error correction. Dynamic range is
limited by maximum receiver input level and the receiver’s
noise floor.
System dynamic range
30 kHz to 300 kHz
100 dB
1,5
300 kHz to 1.3 GHz
110 dB
2
1.3 GHz to 3 GHz
110 dB
3 GHz to 6 GHz
105 dB
Typical measurement uncertainty
3
The following graphs show the typical measurement
uncertainty for the HP 8753E over the full frequency range
using full two-port error correction.
Transmission measurements
Magnitude
Reflection measurements
4
Magnitude
Measurement port characteristics
The following specifications show the residual HP 8753E
system uncertainties for corrected performance after
accuracy enhancement using full two-port error correc-
tion. These characteristics apply for an environmental
temperature of 25 ± 5° C, with less than 1° C deviation
from the calibration temperature.
Frequency range
Corrected
30 kHz-300 kHz5 300 kHz-1.3 GHz
1.3 GHz-3 GHz 3 GHz-6 GHz
Directivity
50 dB
50 dB
47 dB
40 dB
Source match
49 dB
42 dB
36 dB
31 dB
Load match
50 dB
50 dB
47 dB
40 dB
Reflection tracking ±0.005 dB
±0.009 dB
±0.019 dB
±0.070 dB
Transmission tracking ±0.014 dB
±0.013 dB
±0.026 dB
±0.065 dB
Phase
Phase