Atec Agilent-8753E User Manual
Page 10
10
HP 8753E Option 011 specifications
1.
At 25˚ C ±5˚ C, relative to 0 dBm output power for the HP 8753E, +10 dBm output power
for the HP 8753E Option 011.
2.
Typical below 300 kHz.
3.
16 MHz to 3 GHz.
4.
16 MHz to 2 GHz.
5. For HP 8753D Option 011 and Option 006, linearity is specified for the ranges of
(–5 to +13 dBm) and (+13 to +18 dBm).
6.
Typical
Test port output characteristics
Frequency characteristics
Range
300 kHz to 3 GHz
30 kHz to 6 GHz (with Option 006)
Resolution
1 Hz
Stability
typically ±7.5 ppm 0° to 55° C
typically ±3 ppm/year
With Option 1D5
typically ±0.05 ppm 0° to 55° C
typically ±0.5 ppm/year
Accuracy
±10 ppm at 25° C ±5° C
Power range
–5 to +20 dBm –5 to +18 dBm
(with Option 006)
Resolution
0.05 dB
Level accuracy
1,2
± 1.0 dB
Level linearity
1,2,5
(–5 dBm to +15 dBm) ±0.25 dB
(15 dBm to 20 dBm) ±0.5 dB
Impedance
50
Ω; typically
≥16 dB RL (<1.38 SWR) to 3 GHz
2
≥14 dB RL (<1.50 SWR) to 6 GHz
Spectral purity
2nd harmonic
3
<–25 dBc at 20 dBm
<–40 dBc at 10 dBm (typical)
<–50 dBc at 0 dBm (typical)
3rd harmonic
4
<–25 dBc at 20 dBm
<–40 dBc at 10 dBm (typical)
<–50 dBc at 0 dBm (typical)
Nonharmonic spurious
Mixer related
<–30 dBc at 20 dBm (typical)
<–55 dBc at 0 dBm (typical)
Test port input characteristics
Frequency range
300 kHz to 3 GHz
30 kHz to 6 GHz (with Option 006)
Average noise level
2
–90 dBm (3 kHz BW, 50 kHz to 3 GHz)
–110 dBm (10 Hz BW, 50 kHz to 3 GHz)
–120 dBm (10 Hz BW, 50 kHz to 3 GHz)
(typical)
–85 dBm (3 kHz BW, 3 to 6 GHz)
–105 dBm (10 Hz BW, 3 to 6 GHz)
–115 dBm (10 Hz BW, 3 to 6 GHz)
(typical)
Maximum input level
0 dBm
Damage level
20 dBm or 35 VDC
Impedance: 50 ohms
≥10 dB RL, 30 kHz to 50 kHz
(Option 006 only)
2
≥20 dB RL, 50 kHz to 300 kHz
(Option 006 only)
2
≥23 dB RL, 300 kHz to 1.3 GHz
≥20 dB RL, 1.3 GHz to 3 GHz
≥7 dB RL, 3 GHz to 6 GHz
(Option 006 only)
6
Frequency response
± 1.0 dB, 300 kHz to 3 GHz
(25° + 5° C)
± 2.0 dB, 3 GHz to 6 GHz
Harmonics (Option 002)
2nd harmonic
3
<–15 dBc at 0 dBm
<–30 dBc at –10 dBm (typical)
<–45 dBc at –30 dBm (typical)
3rd harmonic
4
<–30 dBc at 0 dBm
<–50 dBc at –10 dBm (typical)
<–50 dBc at –30 dBm (typical)
Harmonic measurement accuracy
(25° ±5° C)
16 MHz to 3 GHz
±1.5 dB
3 GHz to 6 GHz
±3 dB (with Option 006)
Harmonic measurement dynamic range
–40 dBc (output = –10 dBm,
input <–15 dBm)
Test Set
and DUT
A/D
Frequency
Reference
Synthesizer/
Pulse Gen.
Phase Lock
Receiver
3.8 to
6.8 GHz
3.8 GHz
Source
4 kHz
RF Out
30 kHz
to 3 GHz
1 MHz
Main
Processor
ROM/RAM
I/O
Front
Panel
Display
HP-IB
Fast
Processor
B
A
R
HP 8753E Option 011 block diagram