beautypg.com

Atec Agilent-8753E User Manual

Page 7

background image

7

System performance summary
HP 8753E (75-ohm systems) type-F test ports

1.

90 dB, 30 kHz to 50 kHz.

2.

100 dB, 300 kHz to 16 MHz due to fixed spurs.

3.

These measurement uncertainty curves utilize an RSS model for the contribution of random
errors such as noise, typical connector repeatabilities, and test set switch; with a worst-case
model for the contributions of dynamic accuracy and residual systematic errors.

4.

The graphs shown for reflection measurements apply to either a one-port device or a two-port
device with more than 6 dB insertion loss.

5.

Typical performance.

6.

Typical below 300 kHz.

The following specifications describe the system perfor-
mance of the HP 8753E network analyzer with an inte-
grated 75-ohm S-parameter test configuration. System
hardware includes the following:

Network analyzer

HP 8753E Option 075

Calibration kit

HP 85039B

Test-port cables

HP 11857B

Dynamic range
These specifications apply to transmission measurements
in the 30 kHz to 3 GHz frequency range at 10 Hz IF BW
with full two-port error correction. Dynamic range is limit-
ed by maximum receiver input level and the receivers
noise floor.

System dynamic range

30 kHz to 300 kHz

95 dB

1,6

300 kHz to 1.3 GHz

105 dB

2

1.3 GHz to 3 GHz

105 dB

Typical measurement uncertainty

3

The following graphs show the typical measurement
uncertainty for the HP 8753E over the full frequency range
using full two-port error correction.

Transmission measurements

Magnitude

Reflection measurements

4

Magnitude

Measurement port characteristics
The following specifications show the residual HP 8753E
system uncertainties for uncorrected performance and
after accuracy enhancement using full two-port error
correction. These characteristics apply for an environmen-
tal temperature of 25 ± 5° C, with less than 1° C deviation
from the calibration temperature. Data is shown for
type-F female reflection port and type-F male
transmission port.

Frequency Range

Corrected

6

30 kHz-300 kHz

300 kHz-1.3 GHz

1.3 GHz-3 GHz

Directivity

38 dB

38 dB

32 dB

Source Match

36 dB

36 dB

30 dB

Load Match

38 dB

38 dB

32 dB

Reflection tracking

±0.008 dB

±0.008 dB

±0.032 dB

Transmission tracking ±0.062 dB

±0.035 dB

±0.078 dB

Phase

Phase