Calibration kits, Hp 8753e accessories – Atec Agilent-8753E User Manual
Page 21
Calibration kits
Vector accuracy enhancement procedures require that the
systematic errors of the measurement system be character-
ized by measuring known devices (standards) on the
system over the frequency range of interest. The following
calibration kits contain precision standards in many differ-
ent connector types. Return loss specifications or typical
values are provided where available for the terminations
and adapters.
HP 85031B 7-mm calibration kit
Contains precision 7-mm standards used to calibrate the
HP 8753E for measurement of devices with precision 7 mm
connectors.
Includes
HP part number
7 mm short/open circuit
85031-60001
7 mm 50-ohm term (two each)
00909-60008
Specifications for terminations
DC to 5 GHz: RL
≥ 52 dB
5 to 6 GHz: RL
≥ 46 dB
HP 85032B 50-ohm type-N calibration kit
Contains precision 50-ohm type-N standards used to cali-
brate the HP 8753E and 50-ohm test sets for measurement
of devices with 50-ohm type-N connectors. Precision
phase-matched 7-mm to type-N adapters are included for
accurate measurements of non-insertable devices.
Includes
HP part number
N-male 50-ohm termination
00909-60009
N-female 50-ohm termination
00909-60010
N-male short circuit
85032-60008
N-female short circuit
85032-60009
N-female open circuit
85032-60012
N-male open circuit
85032-60007
7 mm to N-male adapter
(two each)
85054-60009
7 mm to N-female adapter
(two each)
85054-60001
Specifications for terminations
DC to 3 GHz: RL
≥ 49 dB
2 to 3 GHz:
RL
≥ 46 dB
3 to 6 GHz:
RL
≥ 40 dB
Typical adapter characteristics
DC to 6 GHz:
RL
≥ 30 dB
HP 85033D 3.5-mm calibration kit
Contains a set of precision 3.5-mm standards to calibrate
the HP 8753E and 50-ohm test sets for the measurement
of devices with precision 3.5-mm and SMA connectors.
Precision phase-matched 7-mm to 3.5-mm adapters are
included for accurate measurements of non-insertable
devices.
Includes
HP part number
3.5-mm-male 50-ohm termination
85033-60009
3.5-mm-female 50-ohm termination
85033-60010
3.5-mm-female short
85033-60014
3.5-mm-male short
85033-60013
3.5-mm-female open
85033-60012
3.5-mm-male open
85033-60011
7-mm to 3.5-mm female
adapter (two)
1250-1747
7-mm to 3.5-mm male
adapter (two)
1250-1746
Specifications for terminations
DC to 1.3 GHz:
RL
≥ 46 dB
1.3 to 3 GHz:
RL
≥ 44 dB
3 to 6 GHz:
RL
≥ 38 dB
Typical adapter characteristics
DC to 6 GHz:
RL
≥ 34 dB
HP 85036B 75-ohm type-N calibration kit
Contains a set of precision 75-ohm type-N standards to
calibrate the HP 8753E and 75-ohm test sets for measure-
ment of devices with 75-ohm type-N connectors. Precision
phased matched adapters are included for accurate
measurements of non-insertable devices.
Includes
HP part number
N-male 75-ohm termination
00909-60019
N-female 75-ohm termination
00909-60020
N-female 75-ohm short
85036-60011
N-male 75-ohm short
85036-60012
N-female open
85032-20001
N-male open
85032-60007
N-male to N-male 75-ohm adapter
85036-60013
N-female to N-female 75-ohm adapter
85036-60014
N-male to N-female 75-ohm adapter
85036-60015
Specifications for terminations
DC to 2 GHz:
RL
≥ 46 dB
2 to 3 GHz:
RL
≥ 40 dB
21