beautypg.com

Corrected system performance with 3.5mm connectors – Atec Agilent-PNA Series User Manual

Page 8

background image

8

Corrected system performance
with 3.5mm connectors

Applies to PNA Series analyzer, 85093B (3.5mm, 50

Ω)

Electronic calibration (ECal) module, and N6314A test
port cable using full two-port error correction.

Description

Specification (dB)
300 kHz to 1.3 GHz

1.3 GHz to 3 GHz

3 to 6 GHz

6 to 9 GHz

Directivity

50

52

51

45

Source match

45

43

40

37

Load match

47

47

44

39

Reflection tracking

±0.043

±0.043

±0.055

±0.100

Transmission tracking

±0.050

±0.045

±0.085

±0.140

Transmission uncertainty

Reflection uncertainty