7

Corrected system performance
with 3.5 mm connectors

Applies to PNA Series analyzer, 85033E (3.5 mm, 50

Ω)

calibration kit, and N6314A test port cable using full
two-port error correction.

Description

Specification (dB)
300 kHz to 1.3 GHz

1.3 GHz to 3 GHz

3 to 6 GHz

6 to 9 GHz

Directivity

46

44

38

38

Source match

43

40

37

36

Load match

46

44

38

38

Reflection tracking

±0.006

±0.007

±0.009

±0.010

Transmission tracking

±0.010

±0.020

±0.041

±0.046

Transmission uncertainty

Reflection uncertainty