Data transfer time (ms) – Atec Agilent-PNA Series User Manual
Page 17
![background image](/manuals/318365/17/background.png)
17
Data transfer time (ms)
a
Number of points
51
201
401
1601
SCPI over GPIB
(program executed on external PC)
b
32-bit floating point
4
7
13
41
64-bit floating point
7
14
24
81
ASCII
25
98
189
804
SCPI over 10 Mbit/s LAN
(program executed on external PC)
c
32-bit floating point
5
6
8
21
64-bit floating point
5
9
22
38
ASCII
18
53
98
362
SCPI over 100 Mbit/s LAN
(program executed on external PC)
c
32-bit floating point
3
5
6
12
64-bit floating point
4
6
9
20
ASCII
17
51
92
339
SCPI (program executed in the analyzer)
d
32-bit floating point
2
3
4
7
64-bit floating point
4
5
6
15
ASCII
26
99
198
781
COM (program executed in the analyzer)
e
32-bit floating point
g
1
1
1
2
Variant type
h
1
3
4
19
DCOM over 10 Mbits/s LAN
(program executed on external PC)
f
32-bit floating point
g
2
3
5
14
Variant type
h
5
14
26
100
DCOM over 100 Mbits/s LAN
(program executed on external PC)
f
32-bit floating point
g
2
2
2
4
Variant type
h
3
5
9
35
a. Typical performance.
b.
Measured using a VEE 5.0 program running on a 600 MHz HP Kayak, National Instruments
TM
GPIB card. Transferred complex S
11
data, using "CALC:DATA? SDATA".
c.
Measured using a VEE 5.0 program running on a 600 MHz HP Kayak. Transferred complex S
11
data, using "CALC:DATA? SDATA". Speed dependent on LAN
traffic, if connected to network.
d.
Measured using a VEE 5.0 program running inside PNA Series analyzer. Transferred complex S
11
data, using "CALC:DATA? SDATA".
e.
Measured using a Visual Basic 6.0 program running inside PNA Series analyzer. Transferred complex S
11
data.
f.
Measured using a Visual Basic 6.0 program running on a 600 MHz HP Kayak. Transferred complex S
11
data. Speed dependent on LAN traffic, if connected
to network.
g.
Used Iarray transfer (getComplex) for 32-bit floating point.
h.
Used meas.GetData for Variant type.