beautypg.com

Atec Agilent-PNA Series User Manual

Page 6

background image

6

Corrected system performance
with type-N connectors

Applies to PNA series analyzer with Option 015,
85032F (Type-N, 50

Ω) calibration kit, and N6314A test

port cable using full two-port error correction.

Description

Specification (dB)
300 kHz to 1.3 GHz

1.3 GHz to 3 GHz

3 to 6 GHz

6 to 9 GHz

Directivity

49

46

40

38

Source match

41

40

36

35

Load match

49

46

40

38

Reflection tracking

±0.011

±0.021

±0.032

±0.054

Transmission tracking

±0.011

±0.023

±0.050

±0.062

Transmission uncertainty

Reflection uncertainty