Test port input (continued) – Atec Agilent-PNA Series User Manual
Page 11
11
Test port input (continued)
Description
Specification
Supplemental information
Reference level magnitude
Range
±200 dB
Resolution
0.001 dB
Reference level phase
Range
±500°
Resolution
0.01°
Stability magnitude
a
300 kHz to 3 GHz
0.02 dB/°C, typical
3 GHz to 6 GHz
0.04 dB/°C, typical
6 GHz to 9 GHz
0.06 dB/°C, typical
Stability phase
a
300 kHz to 3 GHz
0.2°/°C, typical
3 GHz to 6 GHz
0.3°/°C, typical
6 GHz to 9 GHz
0.6°/°C, typical
Maximum test port input level (Test port 1,2)
300 kHz to 25 MHz
+10 dBm
<0.6 dB compression
25 MHz to 3 GHz
+10 dBm
<0.4 dB compression
3 GHz to 6 GHz
+10 dBm
<0.7 dB compression
6 GHz to 9 GHz
+5 dBm
<0.7 dB compression
Maximum receiver input level (A, B, R1, R2)
300 kHz to 6 GHz
–6 dBm, typical
6 GHz to 9 GHz
–11 dBm, typical
Maximum coupler input level (option 015)
300 kHz to 9 GHz
+33 dBm, typical
Reference input level (R1, R2)
b
300 kHz to 9 GHz
–10 to –35 dBm, typical
Damage input level
Test port 1, 2
+30 dBm or ± 30 VDC, typical
R1, R2 IN
+15 dBm or ± 15 VDC, typical
A, B IN (standard)
+15 dBm or ± 15 VDC, typical
A, B IN (option 015)
+15 dBm or 0 VDC, typical
Coupler IN (option 015)
+36 dBm or ± 25 VDC, typical
a. Stability is defined as a ratio measurement measured at the test port.
b. Input level to maintain phase-lock.