Test port input (continued)

Description

Specification

Supplemental information

Reference level magnitude

Range

±200 dB

Resolution

0.001 dB

Reference level phase

Range

±500°

Resolution

0.01°

Stability magnitude

a

300 kHz to 3 GHz

0.02 dB/°C, typical

3 GHz to 6 GHz

0.04 dB/°C, typical

6 GHz to 9 GHz

0.06 dB/°C, typical

Stability phase

a

300 kHz to 3 GHz

0.2°/°C, typical

3 GHz to 6 GHz

0.3°/°C, typical

6 GHz to 9 GHz

0.6°/°C, typical

Maximum test port input level (Test port 1,2)

300 kHz to 25 MHz

+10 dBm

<0.6 dB compression

25 MHz to 3 GHz

+10 dBm

<0.4 dB compression

3 GHz to 6 GHz

+10 dBm

<0.7 dB compression

6 GHz to 9 GHz

+5 dBm

<0.7 dB compression

Maximum receiver input level (A, B, R1, R2)

300 kHz to 6 GHz

–6 dBm, typical

6 GHz to 9 GHz

–11 dBm, typical

Maximum coupler input level (option 015)

300 kHz to 9 GHz

+33 dBm, typical

Reference input level (R1, R2)

b

300 kHz to 9 GHz

–10 to –35 dBm, typical

Damage input level

Test port 1, 2

+30 dBm or ± 30 VDC, typical

R1, R2 IN

+15 dBm or ± 15 VDC, typical

A, B IN (standard)

+15 dBm or ± 15 VDC, typical

A, B IN (option 015)

+15 dBm or 0 VDC, typical

Coupler IN (option 015)

+36 dBm or ± 25 VDC, typical