Measurement throughput summary, Cycle time vs. if bandwidth, Cycle time vs. number of points – Atec Agilent-PNA Series User Manual
Page 16: Cycle time
![background image](/manuals/318365/16/background.png)
16
Measurement throughput summary
Cycle time vs. IF bandwidth
a
Instrument state: preset condition, 201 points, CF = 1 GHz,
Span = 100 MHz, correction off, display off. Add 21 ms
for display on. Cycle time includes sweep and re-trace
time.
Cycle time vs. number of points
a
Instrument state: preset condition, 35 kHz IF bandwidth,
CF = 1 GHz, Span = 100 MHz, correction off, display off.
Add 21 ms for display on. Cycle time includes sweep
and re-trace time.
IF bandwidth (Hz)
Cycle time (ms)
40,000
8
35,000
9
30,000
11
20,000
13
10,000
28
7,000
36
5,000
48
3,000
72
1,000
196
300
620
100
1875
30
8062
10
17877
Number of points Cycle time (ms)
3
4
11
4
51
5
101
6
201
9
401 16
801 29
1601 52
Cycle time
a,b
(ms)
Number of points
51
201
401 1601
Start 800 MHz, Stop 1000 MHz, 35 kHz IF bandwidth
Uncorrected, 1-port cal
6
10
17
53
2-port cal
18
27
39
113
Start 300 kHz, Stop 3 GHz, 35 kHz IF bandwidth
Uncorrected, 1-port cal
32
43
52
93
2-port cal
73
97
117
201
Start 300 kHz, Stop 6 GHz, 35 kHz IF bandwidth
Uncorrected, 1-port cal
40
50
57
98
2-port cal
88
109
125
210
Start 300 kHz, Stop 9 GHz, 35 kHz IF bandwidth
Uncorrected, 1-port cal
45
55
61
99
2-port cal
99
119
133
212
Time Domainc (increase over uncorrected sweep time)
Conversions
<1
<1
4
13
Gating <1
<1
4
17
a. Typical performance.
b. Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with DISPLAY:ENABLE OFF. Add 21 ms for display on.
Data for one trace (S11) measurement.
c. Option 010 only. Analyzer display turned off with DISPLAY:ENABLE OFF. Add 21 ms for display on.