beautypg.com

Measurement throughput summary, Cycle time vs. if bandwidth, Cycle time vs. number of points – Atec Agilent-PNA Series User Manual

Page 16: Cycle time

background image

16

Measurement throughput summary

Cycle time vs. IF bandwidth

a

Instrument state: preset condition, 201 points, CF = 1 GHz,
Span = 100 MHz, correction off, display off. Add 21 ms
for display on. Cycle time includes sweep and re-trace
time.

Cycle time vs. number of points

a

Instrument state: preset condition, 35 kHz IF bandwidth,
CF = 1 GHz, Span = 100 MHz, correction off, display off.
Add 21 ms for display on. Cycle time includes sweep
and re-trace time.

IF bandwidth (Hz)

Cycle time (ms)

40,000

8

35,000

9

30,000

11

20,000

13

10,000

28

7,000

36

5,000

48

3,000

72

1,000

196

300

620

100

1875

30

8062

10

17877

Number of points Cycle time (ms)

3

4

11

4

51

5

101

6

201

9

401 16
801 29

1601 52

Cycle time

a,b

(ms)

Number of points

51

201

401 1601

Start 800 MHz, Stop 1000 MHz, 35 kHz IF bandwidth
Uncorrected, 1-port cal

6

10

17

53

2-port cal

18

27

39

113

Start 300 kHz, Stop 3 GHz, 35 kHz IF bandwidth
Uncorrected, 1-port cal

32

43

52

93

2-port cal

73

97

117

201

Start 300 kHz, Stop 6 GHz, 35 kHz IF bandwidth
Uncorrected, 1-port cal

40

50

57

98

2-port cal

88

109

125

210

Start 300 kHz, Stop 9 GHz, 35 kHz IF bandwidth
Uncorrected, 1-port cal

45

55

61

99

2-port cal

99

119

133

212

Time Domainc (increase over uncorrected sweep time)
Conversions

<1

<1

4

13

Gating <1

<1

4

17

a. Typical performance.
b. Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with DISPLAY:ENABLE OFF. Add 21 ms for display on.

Data for one trace (S11) measurement.

c. Option 010 only. Analyzer display turned off with DISPLAY:ENABLE OFF. Add 21 ms for display on.