beautypg.com

System performance (continued) – Atec Agilent-8720D User Manual

Page 8

background image

8

Agilent 8722D with Option 007 or 085

Calibration kit: Agilent 85052A 2.4 mm with sliding loads
Cables: Agilent 85131F 2.4 mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

Corrected measurement port characteristics

Frequency Range (GHz)

0.05 to 2

2 to 8

8 to 20

20 to 40

Directivity

42 dB

42 dB

42 dB

38 dB

Source Match

40 dB

35 dB

34 dB

31 dB

Load Match

41 dB

48 dB

37 dB

35 dB

Reflection Tracking

±0.011 dB

±0.037 dB

±0.039 dB

±0.047 dB

Transmission Tracking

±0.019 dB

±0.054 dB

±0.082 dB

±0.145 dB

System performance (continued)

Measurement uncertainty
Reflection measurements

Transmission measurements

Magnitude

Magnitude

Phase

Phase