beautypg.com

System performance – Atec Agilent-8720D User Manual

Page 3

background image

3

Agilent 8719D, 8720D with 3.5 mm test ports

Standard, Options 400, 012, 089, or any combination of
these options

Calibration kit: Agilent 85052B, 3.5 mm with sliding loads
Cables: Agilent 85131F 3.5 mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

System dynamic range

0.05 to 2 GHz

100 dB

1

2 to 8 GHz

100 dB

8 to 20 GHz

100 dB

1. Rolls off below 840 MHz to 77 dB at 50 MHz

Corrected measurement port characteristics

Frequency Range (GHz)

0.05 to 0.5

0.5 to 2

2 to 8

8 to 20

Directivity

48 dB

48 dB

44 dB

44 dB

Source Match

40 dB

40 dB

33 dB

31 dB

Load Match

48 dB

48 dB

44 dB

44 dB

Reflection Tracking

±0.006 dB

±0.006 dB

±0.006 dB

±0.008 dB

Transmission Tracking

±0.017 dB

±0.018 dB

±0.066 dB

±0.099 dB

Maximum output power
+5 dBm

System performance

Measurement uncertainty
Reflection measurements

Transmission measurements

Magnitude

Magnitude

Phase

Phase