7

Agilent 8719D or 8720D with Option 007 or 085

Calibration kit: Agilent 85052B 3.5 mm with sliding loads
Cables: Agilent 85131F 3.5 mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

Corrected measurement port characteristics

Frequency Range (GHz)

0.05 to 0.5

0.5 to 2

2 to 8

8 to 20

Directivity

48 dB

48 dB

44 dB

44 dB

Source Match

40 dB

39 dB

32 dB

30 dB

Load Match

48 dB

45 dB

38 dB

37 dB

Reflection Tracking

±0.006 dB

±0.010 dB

±0.030 dB

±0.031 dB

Transmission Tracking

±0.011 dB

±0.016 dB

±0.070 dB

±0.122 dB

Measurement uncertainty
Reflection measurements

Transmission measurements

Magnitude

Magnitude

Phase

Phase