Option 089

Option 089 adds frequency offset mode, allowing
the receiver to be offset from the source frequency.
This allows direct conversion loss measurement of
mixers without need for a reference mixer. RF and
IF frequencies must be within the specified operat-
ing range of the instrument. This test set modifica-
tion adds an internal reference channel switch and
deletes the reference channel output. Firmware
guides the user through test setup. When not in
frequency offset mode, the system operates as a
standard system and meets standard instrument
specifications.

Supplemental characteristics (Option 089)

8719D/8720D

8722D

Reference (R) Input Level
Minimum

-34 dBm

-34 dBm

Maximum (for 0.1 dB compression)

-7 dBm

-12 dBm

Maximum (damage level)

0 dBm

0 dBm

Note: To utilize full instrument receiver dynamic range, measure test signal
at port 2. This configuration requires a power splitter and reference mixer
to provide a phase lock signal to the R channel input.

System performance (continued)

Analyzer’s guided test setup display

Option 089 test set block diagram

Mechanical
transfer switch

R channel
input

LO source

Source

Samplers

Reference

switch

Mixer

under test

A

R

RF

IF

LO

B

Filter