Agilent 8719D Option 400
Agilent 8720D Option 400
Agilent 8722D Option 400
(with 3.5 mm test ports using TRL)
Includes instruments with Options 012 and/or 089
Calibration kit: Agilent 85052C, 3.5 mm for TRL
Cables: Agilent 85131F 3.5 mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)
System dynamic range
Frequency Range
8719D/8720D
8722D
0.05 to 2 GHz
100 dB
1
93 dB
2
2 to 8 GHz
100 dB
93 dB
8 to 20 GHz
100 dB
91 dB
20 to 40 GHz
—
80 dB
1. Rolls off below 840 MHz to 77 dB at 50 MHz
2. Rolls off below 840 MHz to 67 dB at 50 MHz
Corrected measurement port characteristics
8719D/8720D Option 400
Frequency Range (GHz)
0.05 to 0.5
0.5 to 2
2 to 8
8 to 20
Directivity
48 dB
48 dB
50 dB
50 dB
Source Match
40 dB
40 dB
50 dB
50 dB
Load Match
48 dB
48 dB
50 dB
50 dB
Reflection Tracking
±0.006 dB
±0.006 dB ±0.005 dB ±0.005 dB
Transmission Tracking
±0.020 dB
±0.026 dB ±0.015 dB ±0.019 dB
8722D Option 400
Frequency Range (GHz)
0.05 to 2
2 to 8
8 to 20
20 to 26.5
Directivity
48 dB
50 dB
50 dB
50 dB
Source Match
40 dB
50 dB
50 dB
50 dB
Load Match
48 dB
50 dB
50 dB
50 dB
Reflection Tracking
±0.006 dB
±0.005 dB
±0.005 dB ±0.005 dB
Transmission Tracking
±0.017 dB
±0.013 dB
±0.016 dB ±0.023 dB
Maximum output power
8719D/8720D: +5 dBm
8722D (0.05 to 20 GHz): –5 dBm
(20 to 40 GHz): –10 dBm
Measurement uncertainty
Reflection measurements
Transmission measurements
Magnitude
Phase
Magnitude
Phase