beautypg.com

System performance (continued) – Atec Agilent-8720D User Manual

Page 6

background image

6

Agilent 8719D, 8720D, 8722D Option 007
Agilent 8719D, 8720D, 8722D Option 085

Includes instruments with Options 012 and/or 089

Option 007 replaces the standard solid-state trans-
fer switch with a mechanical switch to provide
higher output power.

Option 085 adds internally controlled 0 to 55 dB
step attenuators (5 dB steps) in the receiver path
of both ports, an RF loop that allows the addition
of an amplifier before the transfer switch, and RF
loops after the switch that allow insertion of isola-
tors, required for measurements above 1 watt. An
internal reference channel switch is added and
internal bias tees are deleted. This system is capa-
ble of full two-port calibrated measurements to 20
watts. Measurements up to 100 watts may be possi-
ble using specific configurations. Option 085 is not
compatible with Option 400.

System dynamic r

1

ange

Option 007

Option 085

Frequency Range

8719D/20D 8722D

8719D/20D 8722D

0.05 GHz

82 dB

72 dB

77 dB

67 dB

0.05 to 2 GHz

1

105 dB

98 dB

100 dB

93 dB

2 to 8 GHz

105 dB

98 dB

100 dB

93 dB

8 to 20 GHz

105 dB

96 dB

100 dB

91 dB

20 to 40 GHz

85 dB

77 dB

1. Rolls off below 840 MHz to specified value at 50 MHz

Maximum output power

Option 007

Option 085

2

8719D/8720D

+10 dBm

+5 dBm

8722D (0.05 to 20 GHz)

0 dBm

–5 dBm

8722D (20 to 40 GHz)

–5 dBm

–10 dBm

2. With jumper cable installed between RF out and RF in ports,

i.e. no external amplification.

Supplemental characteristics (Option 085)
Maximum R-channel input level: 0 dBm
Minimum R-channel input level: –34 dBm
Maximum RF port input: +43 dBm
Attenuators: 55 dB maximum, 5 dB steps
Maximum test port power (no isolators): +30 dBm
Maximum test port power (with high power isolators): +43 dBm

System performance (continued)

Option 085 block diagram and example
high power measurement setup

Mechanical
transfer switch

Source

Samplers

Reference

switch

Customer-supplied

booster amplifier and coupler

RF out

RF in +43 dBm max

A

R

B

–55 dB

–55 dB

Customer-supplied isolation
(for output above +30 dBm)

Ref input

0 dBm max

Amplifier
under test

+43 dBm max output
(20 watts)

+43 dBm max input