Frequency Characteristics
Range

8719D

0.05 to 13.51 GHz

S-1

8720D

0.05 to 20.05 GHz

S-1

8722D

0.05 to 40 GHz

S-1

Accuracy (at 23 °C ± 3 °C)

±10 ppm

S-1

Stability

0 °C to 55 °C

±7.5 ppm

C

Option 1D5

±0.05 ppm

C

Per year (aging)

±3 ppm

C

Option 1D5

±0.5 ppm

C

Resolution

1 Hz

S-3

System Characteristics
Maximum Input Level

Damage Level (test port)

30 dBm

C

Reference (F) Input Level (Option 089)

Maximum

8719D/8720D

–7 dBm

C

8722D

–12 dBm

C

Minimum

8719D/8720D/8722D

–34 dBm

C

High Level Trace Noise

2

Magnitude (zero-peak)

0.05 to 13.5 GHz

0.03 dB

C

13.5 to 20 GHz

0.04 dB

C

20 to 40 GHz

0.15 dB

C

Phase (zero-peak)

0.05 to 13.5 GHz

0.3°

C

13.5 to 20 GHz

0.4°

C

20 to 40 GHz

1.5°

C

Spectral Purity Characteristics
Harmonics at maximum output level

<–15 dBc

C

Phase Noise

to 60 kHz from carrier at 2 GHz

<–55 dBc

C

to 60 kHz from carrier at 20 GHz

<–35 dBc

C

Nonharmonic Spurious Signals

at 100 kHz offset

<–40 dBc

C

at 200 kHz offset

<–45 dBc

C

at >200 kHz offset

<–65 dBc

C

Description

Specification

Code

Power Characteristics
Power Range

8719D/8720D

75 dB

C

8722D

0.05 to 20 GHz

70 dB

C

20 to 40 GHz

65 dB

C

Maximum Output Power

8719D/8720D (Standard, Options. 085, 400)

+5 dBm

C

8719D/8720D (Option 007)

+10 dBm

C

8722D (Standard, Options 085, 400)

0.05 to 20 GHz

–5 dBm

C

20 to 40 GHz

–10 dBm

C

8722D (Option 007)

0.05 to 20 GHz

0 dBm

C

20 to 40 GHz

–5 dBm

C

Resolution

0.01 dB

S-3

Flatness (at 5 dB below maximum output power)

8719D/8720D

±2 dB

S-1

8722D

±3 dB

S-1

Power Sweep Range

8719D

20 dB

S-3

8720D

20 dB

S-3

8722D

15 dB

S-3

Power Linearity

±5 dB from reference

±0.35 dB

S-1

–10 dB from reference

±0.6 dB

S-1

+10 dB

1

from reference

±1 dB

S-1

Test Reference Power
8719D/8720D (Standard, Options 085, 400) –5 dBm

S-3

8719D/8720D (Option 007)

0 dBm

S-3

8722D (Standard, Options 085, 400)

–10 dBm

S-3

8722D (Option 007)

Specifications and characteristics

Frequency Range (GHz)

Description

0.05

5

0.05 to 2

5

2 to 8

8 to 20

20 to 40

Code

System Characteristics
Dynamic Range

3

8719D/8720D
(Standard, Options 085, 400)

77 dB

100 dB

100 dB

100 dB

S-1

8719D/8720D (Option 007)

82 dB

105 dB

105 dB

105 dB

S-1

8722D (Standard

4

, Options 085, 400)

67 dB

93 dB

93 dB

91 dB

80 dB

6

S-1

8722D (Option 007)

72 dB

98 dB

98 dB

96 dB

85 dB

S-1

Receiver 0.1 dB Compression Input Level

7

20 dBm

16 dBm

15 dBm

8 dBm

3 dBm

C

1. Does not apply to 8722D.
2. Trace noise is defined as variation of a high signal level trace due to noise. The

value given represents a noise variation that is three standard deviations away
from the trace’s mean value as measured in a 3 kHz IF bandwidth in th fast
sweep mode (STEP SWP OFF).

3. The dynamic range specifications apply to transmission measurements using

10 Hz IF BW and response and isolation correction or full two-port correction.
Dynamic range is limited by the maximum test port power and the receiver’s
noise floor.

4. With 85133E flexible cable on test port.
5. Rolls off below 840 MHz to specified value at 50 MHz.
6. 77 dB for Option 085.
7. For Option 012 direct sampler input compression levels, refer to page 9.

S-1: This performance parameter is verifiable using performance tests documented in

the service manual.

S-2: Due to limitations on available industry standards, the guaranteed performance

of the instrument cannot be verified outside the factory. Field procedures can
verify performance with a confidence prescribed by available standards.

S-3: These specifications are generally digital functions or are mathematically

derived from tested specifications, and can therefore be verified by functional
pass/fail testing.

C:

Non-warranted performance characteristics are intended to provide information
useful in applying the instrument. Performance characteristics are representa-
tive of most instruments, though not necessarily tested in each unit. Not field
tested.