Xylem pHotoFlex/pHotoFlex Turb User Manual
Page 74

Operation
pHotoFlex / pHotoFlex Turb
74
ba75491d04
07/2006
Output format, CSV
Data
Description
1
Date
Date of storing
2
Time
Time of storing
3
ID
adjusted ID
4
Program number
only for the measured parameter,
Concentration
5
Citation form
only for the measured parameter,
Concentration
6
AR/
BV
! AutoRead (measuring mode, pH & ORP)
! Blank value (measured parameter,
Concentration)
7
Dilution
only for the measured parameter,
Concentration
8
Measured value
! Measured value or
! Upper/lower measuring range limit
(only with measured value status, OFL/
UFL)
9
Unit of 8
! Unit of the measured value or
! Designation of dimensionless measured
values, e.g.
10
Measured value
status of 8
! VALID: Measured value valid
! INVALID: Measured value invalid
! UFL: Measured value below the lower
measuring range limit
! OFL: Measured value above the upper
measuring range limit
11
Temperature val-
ue / wavelength
of the measure-
ment
! Temperature (measuring mode, pH &
ORP)
! Wavelength (measured parameter,
Absorbance / % Transmission)
15.01.05;15:06:49;1;8;O2;[BW];[9 + 1];6.5;mmol/l;VALID;
15.01.05;15:09:52;0;;;[AR];;7.00;
11 12 13
1
2
3 4 5
6
7
8
9
10