4 calibration, Calibration – Xylem pHotoFlex/pHotoFlex Turb User Manual
Page 65

pHotoFlex / pHotoFlex Turb
Operation
65
ba75491d04
07/2006
4.7.4
Calibration
When to calibrate?
! Regularly every 90 days
! With a temperature change
Calibration procedures
and calibration
standards
For the menu-guided three-point calibration you need the following
three calibration standards in the mentioned order:
Preparing calibration
Perform the following preparatory activities when you want to calibrate:
Caution
Never open the cells with the calibration standards.
Carrying out calibration
Standard no.
NTU/FNU
1
1000
2
10,0
3
0,02
1
Keep the cells with the required calibration standards ready
and mark them as necessary (see page 62).
2
Clean the cell (see section 5.2.2).
3
Insert the cell (see section 4.2).
1
Press the
ity measuring mode is selected.
2
Press the
The menu-guided calibration begins.
Follow the instructions on the display.
3
Insert the cell with the displayed calibration standard (here e.g.
1000 NTU/FNU) in the cell shaft (see section 4.2).
i Insert standard
1000 FNU/NTU
i Press and hold
Turbidity \ Calibration