Calibration

When to calibrate?

! Regularly every 90 days
! With a temperature change

Calibration procedures

and calibration

standards

For the menu-guided three-point calibration you need the following
three calibration standards in the mentioned order:

Preparing calibration

Perform the following preparatory activities when you want to calibrate:

Caution

Never open the cells with the calibration standards.

Carrying out calibration

Standard no.

NTU/FNU

1

1000

2

10,0

3

0,02

1

Keep the cells with the required calibration standards ready
and mark them as necessary (see page 62).

2

Clean the cell (see section 5.2.2).

3

Insert the cell (see section 4.2).

1

Press the key (long pressure) repeatedly until the Turbid-
ity
measuring mode is selected.

2

Press the key.

The menu-guided calibration begins.
Follow the instructions on the display.

3

Insert the cell with the displayed calibration standard (here e.g.
1000 NTU/FNU) in the cell shaft (see section 4.2).

i Insert standard

1000 FNU/NTU

i Press and hold
i Align sample

Turbidity \ Calibration