5 measuring the absorbance/transmission, Measuring the absorbance/transmission – Xylem pHotoFlex/pHotoFlex Turb User Manual
Page 45

pHotoFlex / pHotoFlex Turb
Operation
45
ba75491d04
07/2006
4.5.5
Measuring the absorbance/transmission
Note
The transmission measurement is not described separately in the fol-
lowing example as it operates in exactly the same way as the absor-
bance measurement. The result of the measurement is displayed in %
Transmission.
8
Confirm the result with
The blank value measurement is completed.
The meter is ready to measure.
or:
Discard the result with
Subsequently, carry out a new blank value measurement.
1
Press the
tometry measuring mode is selected.
2
Press the
sured variable, Absorbance or % Transmission is selected.
3
Select the cell diameter with <
▲> <▼> and confirm with
4
Select the wavelength with <
▲> <▼> and confirm with
i 2: A5/25 MC
i
Cell = 16 mm
i Absorbance = 0.301
i Blank value mea-
surement
successful!
Measure blank value
Photometry \ Absorbance
i Select cell
with
▲
▼
16 mm
610 nm
01.02.05 15:12