Xylem pHotoFlex/pHotoFlex Turb User Manual
Page 64

Operation
pHotoFlex / pHotoFlex Turb
64
ba75491d04
07/2006
Note
To keep the drift as low as possible, the time for aligning the cell while
pressing and holding the
onds. After this time, the meter automatically starts measuring or cali-
brating.
Display if the
measuring range is
exceeded
If the measured value is outside the measuring range of the
pHotoFlex Turb, it is indicated on the display:
7
Align the cell:
! Marked cell
– Align the marking on the cell cap with the marking on the
cell shaft.
– Press and for a short time hold the
key until the measured value is displayed.
! Unmarked cell (see page 62)
– Press and hold the
– Slowly and in small steps turn the cell by one complete
rotation (by 360 °). After each step wait for a short time
until the displayed measured value is stable.
– Turn the cell back to the position with the lowest mea-
sured value.
8
Release the
Measurement starts. The measured value is displayed.
9
Repeat the steps 2 to 8 for further samples.
Turbidity
157.0
FNU
NTU
01.02.05 15:12
Turbidity
> 1100
FNU
NTU
01.02.05 15:12