beautypg.com

Xylem pHotoFlex/pHotoFlex Turb User Manual

Page 44

background image

Operation

pHotoFlex / pHotoFlex Turb

44

ba75491d04

07/2006

Measuring the blank

value

Note
The following measurement of the blank value applies only to the se-
lected program. If no program is selected, the message

i No program

selected. appears on the display.

1

Press the key (long pressure) repeatedly until the Pho-
tometry
measuring mode is selected.

2

Press the key (short pressure) repeatedly until the mea-
sured variable, Concentration is selected.

3

Select a program with as necessary.

4

Open the adjustment menu with .

5

Using <

> <> and , select and start the

Measure blank value function.

The menu-guided blank value measurement starts.
Follow the instructions on the display.

6

Insert a cell with blank sample (see section 4.2).

7

Start the measurement of the blank value with TER>.

After measuring, the result of the blank value measurement is
displayed and stored.
The result is displayed as

i Blank value measurement suc-

cessful! or

i Blank value measurement erroneous!

Zero adjust.

Measure blank value

Delete blank value

Photometry \ Adjustment

i 2: A5/25 MC
i

Cell = 16 mm

i Insert blank sample

Start measurement

Measure blank value