Xylem pHotoFlex/pHotoFlex Turb User Manual
Page 59

pHotoFlex / pHotoFlex Turb
Operation
59
ba75491d04
07/2006
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the asymmetry of the elec-
trode.
5
Using <
▲> <▼>, select Continue and press
TER>. The buffer is measured.
The measured value is checked for stability (AutoRead).
6
Wait for the end of the AutoRead measurement or accept the
calibration value with
The calibration display for the next buffer appears.
7
For single-point calibration, select Exit with one point calibra-
tion with <
▲> <▼> and confirm with
The calibration is completed as a single-point calibration.
The new calibration values are displayed as an informative
message (
i
).
You have the following options:
! Accept the new calibration values with
Subsequently, the calibration record is displayed and output
to the interface at the same time.
! To switch to the measured value display without accepting
the new calibration values, press
i Buffer value = 7.00
i U = 3 mV
i Temperature =
24.8 °C
Terminate AutoRead
pH & ORP \ Calibration
i Buffer recognition TEC
i Immerse sensor in buffer 2
Exit with one point calibration
Continue
pH & ORP \ Calibration