4 blank value (reagent blank value), Blank value (reagent blank value) – Xylem pHotoFlex/pHotoFlex Turb User Manual
Page 43

pHotoFlex / pHotoFlex Turb
Operation
43
ba75491d04
07/2006
4.5.4
Blank value (reagent blank value)
A blank value is required for every concentration measurement.
For some programs (methods) for concentration measurement, the
blank values are already stored in the photometer. They are used au-
tomatically. For all other programs, the blank value has to be deter-
mined separately before the first measurement.
Each stored reagent blank value can be replaced by a blank value de-
termined by the user.
Note
You will find more information on blank values in the photometry anal-
ysis manual.
A blank value is always stored for the program that has just been called
up. It remains stored until it is erased (menu item, Delete blank value)
or overwritten.
The Reset function erases all blank values measured by the user and
restores the blank values stored in the factory.
If a blank value measured by the user is stored for a program, this blank
value is used for measurement. The usage of the blank value mea-
sured by the user is documented and also indicated in the measured
value display.
4
Insert the cell (see section 4.2).
5
Start the measurement with
Measurement is started. The result is displayed.
Photometry \ Concentration
i Insert sample
i Start measurement
with
1: A5/25 MC
NH4-N
16 mm
0.20 - 6.51 mg/l
01.02.05 15:12
Photometry \ Concentration
[BV]
0.74
mg/l
1: A5/25 MC
NH4-N
16 mm
0.20 - 6.51 mg/l
01.02.05 15:12
A blank value measured by
the user is used