Xylem pHotoFlex/pHotoFlex Turb User Manual
Page 56

Operation
pHotoFlex / pHotoFlex Turb
56
ba75491d04
07/2006
Calibration points
Calibration can be performed using one, two or three buffer solutions in
any order (single-point-, two-point or three-point calibration). The meter
determines the following values and calculates the calibration line as
follows:
Note
You can display the slope in the unit, mV/pH or % (see section 4.6.4).
AutoRead
The calibration procedure automatically activates the AutoRead func-
tion.
The current AutoRead measurement can be terminated at any time
(accepting the current value).
Calibration record
When finishing a calibration, the new calibration values are displayed
as an informative message (
i
symbol) first. Then you can decide
whether you want to take over these values of the new calibration or
whether you want to continue measuring with the old calibration data.
After accepting the new calibration values the calibration record is dis-
played.
Displaying and down-
loading calibration data
to interface
You can view the data of the last calibration on the display. Subse-
quently, you can download the displayed calibration data to the inter-
face, e. g. to a printer or PC, with the
The calibration record of the last calibration can be found under the
Configuration / pH & ORP / Calibration / Calibration record menu item.
Determined val-
ues
Displayed calibration data
1-point
Asy
! Asymmetry = Asy
! Slope = Nernst slope
(-59.16 mV/pH at 25 °C)
2-point
Asy
Slp.
! Asymmetry = Asy
! Slope = Slp.
3-point
Asy
Slp.
! Asymmetry = Asy
! Slope = Slp.
The calibration line is calculated
by linear regression.