6 zero adjustment, Zero adjustment – Xylem pHotoFlex/pHotoFlex Turb User Manual
Page 46

Operation
pHotoFlex / pHotoFlex Turb
46
ba75491d04
07/2006
4.5.6
Zero adjustment
The zero adjustment, i. e. measuring and storing the absorbance of a
cell filled with water, is necessary after the meter is switched on.
Additionally, we recommend to carry out a zero adjustment if the ambi-
ent temperature has changed.
Only perform the zero adjustment against distilled water in an optically
perfect cell. The zero adjustment must be performed separately for
each cell type.
5
Clean the cell (see section 5.2.2).
6
Insert the cell (see section 4.2).
7
Start the measurement with
The measurement result is displayed when the measurement
is finished.
1
Press the
tometry measuring mode is selected.
2
Press the
sured variable, Concentration is selected.
3
Press the
The menu for adjustment measurements opens up.
Photometry \ Absorbance
i Insert sample
i Start measurement
with
16 mm
610 nm
01.02.05 15:12
Photometry \ Absorbance
0.532
16 mm
610 nm
01.02.05 15:12