Continuing for three-
point calibration
(Cal. type TEC)
14
Thoroughly rinse the electrode with distilled water.
15
Immerse the electrode in buffer solution 3.
16
If necessary, measure the temperature of buffer 3 manually,
then enter and confirm it with <
▲> <▼> and
in the Set temperature setting.
17
Using <
▲> <▼>, select Continue and press
TER>. The buffer is measured.
The measured value is checked for stability (AutoRead).
18
Wait for the end of the AutoRead measurement or Terminate
AutoRead with
value.
The new calibration values are displayed as an informative
message (
i
).
You have the following options:
! Accept the new calibration values with
Subsequently, the calibration record is displayed and output
to the interface at the same time.
! To switch to the measured value display without accepting
the new calibration values, press
i Buffer value = 4.01
i U = 184 mV
i Temperature =
24.8 °C
Terminate AutoRead
pH & ORP \ Calibration
- Lab 850 BENCHTOP pH METER (73 pages)
- Lab 850, up to serial number 09470314 (61 pages)
- Lab 860 BENCHTOP pH METER (74 pages)
- Lab 870 BENCHTOP pH METER (71 pages)
- Lab 960 (77 pages)
- Lab 970 (71 pages)
- CG 839 (8 pages)
- CG 842 (44 pages)
- CG 843P (72 pages)
- CG 843 (68 pages)
- CG 853P (76 pages)
- CG 853 (74 pages)
- CG 855 (11 pages)
- HandyLab 100 QuickStart (50 pages)
- HandyLab 100 (66 pages)
- HandyLab 200 QuickStart (50 pages)
- HandyLab 200 (52 pages)
- HandyLab 600 QuickStart (50 pages)
- HandyLab 600 (72 pages)
- HandyLab 680 QuickStart (50 pages)
- HandyLab 680 (120 pages)
- HandyLab 700(EX) (128 pages)
- HandyLab 700 (76 pages)
- HandyLab 780 (164 pages)
- handylab LF 11 (44 pages)
- handylab LF 12 (64 pages)
- handylab OX 12 (60 pages)
- handylab multi 12 (82 pages)
- handylab pH 11 (48 pages)
- handylab pH 12 (68 pages)
- handylab pH LF 12 (80 pages)
- IQ SensorNet FDO 70x IQ (SW) (48 pages)
- IQ SensorNet ViSolid 700 IQ (SW) (58 pages)
- MultiLab 4010-2 (112 pages)
- MultiLab 4010-3 (112 pages)
- EBI 40 TK-12 (18 pages)
- TFX 410 (38 pages)
- EBI 20 Serie (20 pages)
- EBI20 TEMPERATURE/HUMIDITY DATALOGGERS (40 pages)
- EBI 20-TF (40 pages)
- TFX 422 (36 pages)
- ST 1000 (42 pages)
- TLC 1598 (2 pages)
- TTX 100/110/120 (28 pages)
- EBI 25 (18 pages)