Appendix b. mjk-field link software, Important notes, A. connect a pc to the oxix – Xylem Oxix Dissolved Oxygen Transmitter User Manual
Page 89: Important notes! a. connect a pc to the oxix

89
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Appendix B. MJK-Field Link Software
The MJK-Field Link software package provides several utilities that will be
described in detail on the following pages:
PC <-> Oxix
•
®
transmitter interconnection
Saving log data
•
Display firmware upgrade
•
Converter firmware upgrade
•
Language installation
•
Important Notes!
1. Current Oxix
®
log data must be saved in a file before transferring and
installing new firmware. See "2. Save Log Data" overleaf.
2. Note down all display and converter settings before upgrading new
firmware.
3. The upgrade procedures below must be carried out in strict alphabetical
order. First procedure A, then B, C, D and finally E.
A. Connect a PC to the Oxix
®
1. Insert the Oxix
®
MJK-Field Link CD-ROM into the PC's disk drive.
The MJK-Field Link opening menu will appear after a short while (see be-
low). If the auto-run function does not execute, locate the
"MJK-Field-Link 840110-007.exe" file (or a similar file) and click "Run".