Sensor installation – Xylem Oxix Dissolved Oxygen Transmitter User Manual

Page 19

background image

19

Dissolved Oxygen Transmitter

GB Oxix Manual 080915

FW: 841013-009, 846001-002, 841513-009

4. Sensor Installation

WARNING!
The Oxix

®

transmitter must not be powered until the transmitter is

correctly mounted and connected!
WARNING!
Connections to the Oxix

®

transmitter must not be disconnected while

the transmitter is powered.
Important!
To ensure safe and optimal functionality of the Oxix

®

transmitter,

the instructions below must be followed closely.

The sensor must be mounted in a location where it is possible to perform

a representative dissolved oxygen sat. measurement.
Fittings and mountings must be capable of carrying the sensor's

weight

(0,77 kg / 1.7 lbs), and vibrations from for example pumps and other
equipment must be avoided.
The system should be installed to facilitate access to the Oxix

®

transmitter

operator panel and sensor without risking personal injury during normal
operation, maintenance, cleaning of sensor, etc.
The Oxix

®

transmitter must be connected to a mains outlet that complies

with national and local regulations.
All cables must be installed, connected and mounted to minimize me-

chanical impact and electrical influence from other installations.
The sensor cable’s maximum tensile load is adjusted to the sensor weight.

The sensor is a electro-mechanical instrument. Be careful not to damage

the sensor lens. The accuracy will deteriorate if foreign material accumu-
lates on it, or it is subject to abrasive cleaning procedures. Only use a soft
cloth for cleaning. NEVER EVER a steel brush!
All

fuses are located on the Oxix

®

converter’s motherboard.