Appendix A. Pop-up and Error Messages
Alarms
Current overrange
Alarm - The mA connection has exceeded
the 4 - 20 mA range. For example 3,5 or
20,5 mA.
Factory reset occurred
Alarm - A factory reset has occured. Some
settings have changed. Check all used set-
tings.
High alarm
Alarm - The high set point has been de-
tected.
Low alarm
Alarm - The low set point has been de-
tected.
mA connection
Alarm - The mA connection is not connect-
ed. If the current-out is used, deactivate mA
Current out to stop this alarm.
Errors
Calibration
Error - 1. Perform calibration (see page
and/or check the units and ranges (see page
).
Eeprom write error
Error - Eeprom write error. If error keeps
popping up, contact MJK.
Measuring start failed
Error - Start failed. Try again.
Measuring timeout
Error - Timeout. Try again.
Missing adjustment
Error - The measured value has not been
entered.
Not in solid mode!
Error - Calibration can only be performed
when in solid mode. Change mode and try
again.