30

GB Oxix Manual 080915

FW: 841013-009, 846001-002, 841513-009

Dissolved Oxygen Transmitter

Sensor and Mounting Kit / Connection Board Terminals

Important!: Make sure the right wiring indication label is fitted below
the terminal block (label indicated by arrow). The label
must clearly be marked Oxix

®

.

Wrong wiring may cause sensor destruction !

Connect the sensor wires to the Oxix

®

converter with the wires coming from

the sensor as shown below (see arrow).

Mains

supply

-mA-

-D02-

-D01-

-DI-

-Display-

+

+

A

B

C2 N02

C NO

N01

PE

N

L

-

-

C1

Line

2 wire

shielded cable.

4 wire

shielded cable.

displa

y unit

displa

y unit

Display unit

Converter #4

Converter #1

Max. four MJK converters with Modbus
can be connected to one display unit:

521274

2,5

Main supply

Supply

Fuse

10-30

V DC

1 A

T

24

V A

C

500 mA

T

115

V A

C

125 mA

T

230

V A

C

63 mA

T

Sensor supply

Fuse 200 mA T

+DC

GND A

B

Red

Black

White

Gr

een

Shield

521275

www.mjk.com

DK: +45 45 56 06 56
NO: +47 69 20 60 70
SE: +46 53 31 77 50
NL: +31 251 672171
USA: +1 847 482 8655
AUS: +61 3 9758 8533

Oxix

®

converter terminals

Terminals

Wires from sensor

Converter terminals

-

not used

Black

GND

White

A

Green

B

Shield

- (shield GND)

Red

+DC