Dissolved oxygen transmitter, Oxix, Converter terminals – Xylem Oxix Dissolved Oxygen Transmitter User Manual
Page 30

30
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Sensor and Mounting Kit / Connection Board Terminals
Important!: Make sure the right wiring indication label is fitted below
the terminal block (label indicated by arrow). The label
must clearly be marked Oxix
®
.
Wrong wiring may cause sensor destruction !
Connect the sensor wires to the Oxix
®
converter with the wires coming from
the sensor as shown below (see arrow).
Mains
supply
-mA-
-D02-
-D01-
-DI-
-Display-
+
+
A
B
C2 N02
C NO
N01
PE
N
L
-
-
C1
Line
2 wire
shielded cable.
4 wire
shielded cable.
displa
y unit
displa
y unit
Display unit
Converter #4
Converter #1
Max. four MJK converters with Modbus
can be connected to one display unit:
521274
2,5
Main supply
Supply
Fuse
10-30
V DC
1 A
T
24
V A
C
500 mA
T
115
V A
C
125 mA
T
230
V A
C
63 mA
T
Sensor supply
Fuse 200 mA T
+DC
GND A
B
Red
Black
White
Gr
een
Shield
521275
www.mjk.com
DK: +45 45 56 06 56
NO: +47 69 20 60 70
SE: +46 53 31 77 50
NL: +31 251 672171
USA: +1 847 482 8655
AUS: +61 3 9758 8533
Oxix
®
converter terminals
Terminals
Wires from sensor
Converter terminals
-
not used
Black
GND
White
A
Green
B
Shield
- (shield GND)
Red
+DC