Startup, Initial checks before power-up, Initial power-up – Xylem Oxix Dissolved Oxygen Transmitter User Manual
Page 37: Initial measurement, Language selection
37
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
8. Startup
Initial Checks before Power-up
Before switching on power the following steps must be checked.
1. The local mains power supply voltage corresponds to the voltage printed
on the identification label of the transmitter.
2. All electrical connections are made in accordance with the electrical con-
nection diagram shown with special emphasis on the sensor connection
described on page
3. All terminal screws are tightened.
4. All cable glands are tightly secured.
5. All grounding connections are made in accordance with the instructions in
Initial Power-up
1. Turn on power.
2. Within 60 seconds the Oxix
®
sensor will be found by the automatic sen-
sor detection program. During this process you might briefly see "SuSix1"
and/or "MagFlux1" on the display. This is normal and to be expected.
Initial Measurement
1. Make sure that the sensor is completely covered by liquid.
2. Turn on power to the transmitter and wait 15 minutes before taking the
initial measurement.
Language Selection
1. The default display language is English. If another language is required,
proceed with step 2.
2. Select "Display Setup" from the Main Menu.
3. Select "Language" from the Display Setup menu and chose the required
).